Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
START: POS is displayed. (If START: NEG is displayed instead, then press any
one of the arrow keys to toggle to the next state—POS).
7
Press Gate & ExtArm or Enter key.
STOP: NEG is displayed.
Press any one of the arrow keys until STOP: TIME is displayed.
8
9
Set the Gate Time to 5 ms by performing the following steps:
a.
b. Press the appropriate arrow keys until TIME:
c.
NOTE
BE SURE to press the Enter key to complete the numeric entry.
The counter will now gate for a duration of 5 ms. The 5 ms interval will begin after
the external edge is detected.
10
Press Stop/Single key to start the measurement.
The measurement is started after the external arm pulse occurs.
2-18
Press Gate & ExtArm or Enter key.
TIME:
.100 seconds (the default time) is displayed.
Press Enter key.
.00500 s is displayed.
Operating Guide