Instructions for Operation and Care of Dickson Paraffin Baths
DICKSON PARAFFIN BATH
MODELS PB-105 & PB-105B
(Treatment Temperature)
Specification:
Overall Dimensions: 29" L x 15" W x 14" D (737 L x 381 W x 356 D)
Inner Tank Dimensions: 25-1/2" L x 10" W x 10" D (648 L x 254 W x 254 D)
Tank Capacity: 48 lbs. (21.77 kgs.)
Electrical Requirements:

DIRECTIONS FOR OPERATING

Remove Plexiglass Slat from inside unit and make sure to
close drain valve underneath unit (PB-104 & PB-105).
Place the paraffin in the bath. Plug the bath into AC outlet.
There is no power switch so the unit will be on once
plugged in. Rotate the timer switch to activate the timed hi-
heat circuit for the full 55 minutes. To rapidly melt the
paraffin it may be desirable to activate the hi-heat circuit a
second time. In any event, when the time switch returns to
"0" the hi-heat circuit is broken and the bath automatically
will begin to operate on the maintenance circuit. Therefore,
after the timer returns to "0", allow the bath to cool until the
treatment temperature of 126°F (52.2°C) is reached and
maintained. The high heat warning light will go off as
treatment temperature is approached.
** CAUTION!
Do not rely on the light alone to
determine whether the Dickson Paraffin Bath is at
operating temperature – ALWAYS CHECK THAT THE
PROVIDED THERMOMETER READS A SAFE
OPERATING TEMPERATURE, 126° to 132°F (52.2° to
55.6°C) BEFORE SUBMERGING ANY APPENDAGES
IN THE WAX!
4
WHITEHALL MANUFACTURING • P .O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A
Phone (800) 782-7706 • (626) 968-6681 • Fax (626) 855-4862 • Web: www.whitehallmfg.com
Tank
High Heat
(Sanitizing)
Timer
Low Heat
Indicator Light
Inlet Voltage 100-120 V
Frequency 50/60 Hz
6.7 A
High Heat (Sanitizing)
Indicator Light
Inlet Voltage 200-240 V
Frequency 50/60 Hz
3.4 A
** ATTENTION ! Ne vous fiez pas seulement au voyant
pour déterminer si le bain de paraffine Dickson est à la
bonne température - TOUJOURS VÉRIFIER QUE LA
TEMPÉRATURE PRISE AVEC LE THERMOMÈTRE
FOURNI EST LA BONNE TEMPÉRATURE DE
FONCTIONNEMENT, soit de 52,2 à 55,6C (126 à 132F)
AVANT D'IMMERGER TOUT APPENDICE DANS LA
CIRE !
Keep the unit connected to the electrical outlet. The
maintenance circuit, independent of the high-heat circuit, is
controlled by a close differential thermo-switch set to
automatically keep the bath liquid at a temperature of
126°F (52.2°C). The DICKSON Bath precisely maintains
this temperature. This insures the quickest possible "glove"
to be formed.
If a higher temperature is required for an individual patient,
turn on the high heat knob until the thermometer shows the
desired temperature. With the high-heat knob switched on
and the bath covered, the temperature will rise 1°F
(0.55°C) per minute.