JB INDUSTRIES DV-22N Cleaning & Repair Instructions

JB INDUSTRIES DV-22N Cleaning & Repair Instructions

Digital micron gauge

Advertisement

Quick Links

Digital Micron Gauge
Cleaning & repair Instructions
for DV-22N, DV-40, & DV-40S
DV-40
DV-22N
DV-40S
JB INDUSTRIES
CONNECT WITH US
WWW.JBIND.COM
800.323.0811
SALES@JBIND.COM
TM

Advertisement

loading

Summary of Contents for JB INDUSTRIES DV-22N

  • Page 1 Digital Micron Gauge Cleaning & repair Instructions for DV-22N, DV-40, & DV-40S DV-40 DV-22N DV-40S JB INDUSTRIES CONNECT WITH US WWW.JBIND.COM 800.323.0811 SALES@JBIND.COM...
  • Page 2: Cleaning The Sensor

    DIGITAL MICRON GAUGE CLEANING & REPAIRS JB INDUSTRIES If your micron gauge is acting erratic, for instance, vacuum readings at atmosphere, readings rising and falling during a steady pull down or holding vacuum, cleaning the sensor may be necessary. Cleaning the sensor...
  • Page 3 DIGITAL MICRON GAUGE CLEANING & REPAIRS JB INDUSTRIES If you have trouble getting an accurate reading after cleaning the sensor, please check your o-ring for damage, wear, or to see if it is missing. Replacing the O-Ring Tools O-ring probe or thin implement & P90009 Replacement O-ring.
  • Page 4 CONNECT WITH US WWW.JBIND.COM JB INDUSTRIES INC. Aurora IL 60505 USA ©2016 JB Industries, Inc.