Pain Relief System PUMP with ONDEMAND BOLUS BUTTON Instructions for Use...
Page 3
ON-Q* Pump with ONDEMAND* Bolus Button Instructions for Use • There is no alarm or alert when flow interruption Important Information occurs, therefore, life-supporting medications whose usage may cause serious injury or death Please read the entire document before operating due to stoppage or under-delivery are not the ON-Q* device.
Page 4
• It is the responsibility of the healthcare provider to set, close tubing clamp. Replace pump if necessary. modify Patient Guidelines provided with the pump • Do not discard the pump and contact Halyard Health for as appropriate for your patients’ clinical status and product return instructions.
Description of Device: (Figure 1) temperature and will decrease approximately 1.4% per 1°F/0.6°C decrease in temperature. The ON-Q* Pump with ONDEMAND* device incorporates a bolus • If refrigerated, allow pump to reach room temperature device. The ONDEMAND* device delivers a continuous infusion before using.
Instructions for Use Caution: Do not fill less than the labeled fill volume or exceed the maximum fill volume. (Table 1) Use Aseptic Technique Table 1: Fill Volume Filling the ON-Q* Pump: (Figure 2) Note: Follow hospital protocols and applicable regulations for Labeled Retained Maximum Fill Vol.
RIGHT WAY Figure 3-A Figure 3-C Figure 3-D Orange Bottom Bolus Refill Indicator Warning: Do not pull the red tab upwards as breakage could occur (Figure 3-B). If red tab is not removed or breaks while removing, Bolus Activation continuous delivery will occur. This delivery may 1.
• Infusion is complete when pump is no longer inflated. • Close clamp, disconnect and dispose of pump according to your institution’s protocol. Note: If pump did not perform as expected do not discard. Contact Halyard Health for product return instructions at: HalyardIrvineProductComplaint@hyh.com or 1-800-448-3569. Technical Specifications Delivery Accuracy: When filled to the labeled volume, basal flow rate accuracy is ±15% and bolus dose accuracy...