Co2meter RAD-0100 Calibration Procedure

Remote co2 storage safety alarm

Advertisement

RAD‐0100 Remote CO2 Storage Safety Alarm Calibration Procedure 
There are 2 methods for calibration: diffusion (recommended) and pump using a standard gas.
For diffusion calibration, we recommend taking the unit outdoors or anywhere it will be exposed
to fresh air. Wait 5 minutes until the CO2 reading stabilizes, start the calibration, then step away
from the meter to insure your exhaled breath will not disrupt the calibration process.

1. Diffusion Calibration

2. Press the Mode Button until the CALI icon
appears flashing.
1. Check that the Sensor Protector is
removed. You can store it in the lower indent
on the body of the unit.
Sensor Protector
Enter Button
Reset Button
Mode Button
3. Press Enter. The CALI icon will show
steady on the LCD.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RAD-0100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Co2meter RAD-0100

  • Page 1 RAD‐0100 Remote CO2 Storage Safety Alarm Calibration Procedure  There are 2 methods for calibration: diffusion (recommended) and pump using a standard gas. For diffusion calibration, we recommend taking the unit outdoors or anywhere it will be exposed to fresh air. Wait 5 minutes until the CO2 reading stabilizes, start the calibration, then step away from the meter to insure your exhaled breath will not disrupt the calibration process.
  • Page 2 4. Press and hold the Mode button until the 5. After 10 minutes the LCD will display CALIBRATING icon flashes (about 10 PASS or FAIL. seconds). Let go. Calibration will begin. FRESH AIR IS REQUIRED. DO NOT BREATHE ON UNIT. 6.
  • Page 3 Contact us: We are here to help! If this calibration guide does not answer your questions - or for more information - please contact: support@co2meter.com (386) 872-7665 (M-F 9:00am–5:00pm EST) (866) 422-2356 fax www.co2meter.com...

Table of Contents