RAD‐0100 Remote CO2 Storage Safety Alarm Calibration Procedure There are 2 methods for calibration: diffusion (recommended) and pump using a standard gas. For diffusion calibration, we recommend taking the unit outdoors or anywhere it will be exposed to fresh air. Wait 5 minutes until the CO2 reading stabilizes, start the calibration, then step away from the meter to insure your exhaled breath will not disrupt the calibration process.
Page 2
4. Press and hold the Mode button until the 5. After 10 minutes the LCD will display CALIBRATING icon flashes (about 10 PASS or FAIL. seconds). Let go. Calibration will begin. FRESH AIR IS REQUIRED. DO NOT BREATHE ON UNIT. 6.