Is a continuous indicator of the current levels of ec, ph and temperature, enabling you to optimise these parameters as your crop progresses through each growing phase (16 pages)
Page 1
Bluelab Guardian Monitor magically measures all 3 critical parameters for successful growth: pH, conductivity AND temperature Care and use guide www.getbluelab.com...
Page 2
WARNING! Keep your pH probe tip wet at all times to avoid permanent damage...
Page 3
Congratulations on purchasing the Bluelab Guardian Monitor The Bluelab Guardian Monitor is a continuous indicator of the current levels of EC, pH and temperature of your nutrient solution, enabling you to optimise these parameters as your crop progresses through each growing phase.
Page 4
2.0 IMPORTANT - Bluelab pH Probe care 3.0 Mounting the Bluelab Guardian Monitor - if wall mounting The Bluelab pH Probe is the only part of the Bluelab Guardian Monitor that should ever require replacing. Select a suitable location that is less than 2 meters from pH probes DO NOT last forever.
Page 5
Storage and use of calibration solutions the 3 LEDs next to the conductivity window. When the EC/CF LED is on, the Guardian Monitor is displaying conductivity in units of EC or CF. If there is a decimal point present in the conductivity display, the units are EC.
Page 6
11.0 Placement of probes 12.0 Set alarms (optional) cont. Both the Bluelab pH Probe and Bluelab Conductivity/Temperature Probe require submersion in the liquid for Turn alarm ON/OFF a measurement to occur. Press the alarm button to change between alarm ON and alarm OFF. When the alarm is ON the alarm LED will glow.
Cleaning the conductivity/temperature probe periodically ensures accurate readings. The conductivity/ temperature probe is cleaned using the Bluelab Conductivity Probe Cleaner, or “Jif” a trade name for a regularly using the following instructions. The storage cap must always be put back on after cleaning.
Solution > 14.0 pH. pH displays ‘ur’ Under range pH. Solution < 0.0 pH. Check pH probe connection. pH probe could be faulty. Guardian Monitor could be wet inside. temp displays ‘or’ Over range temperature. Solution >51 °C / 122 °F.
Page 9
To ensure you receive a long life from your pH probe, please read the instructions provided with it. When the time comes to replace your Bluelab pH Probe all you have to do is order a replacement from your supplier!
Page 10
The instrument is only as accurate as the probe is clean! Bluelab cleaning kits Probe cleaning is one of the most important parts of owning and operating any Bluelab meter, monitor or controller. If the probe is contaminated (dirty) it affects the accuracy of the reading displayed.
Page 11
Fax: +64 7 578 0847 2 years (24 months) from the date of sale to the original purchaser. Email: support@getbluelab.com (This guarantee does not cover the Bluelab pH Probe. The Bluelab pH Probe is covered by a separate 6 month guarantee). Product details...
Page 12
The Bluelab product range all come with a free repair or replacement guarantee for your added benefit. If you need assistance or advice - we’re here to help you. Phone: +64 7 578 0849 Fax: +64 7 578 0847 Email: support@getbluelab.com...
Need help?
Do you have a question about the Guardian Monitor and is the answer not in the manual?
Questions and answers