QUICK GUIDE FOR ASSISTANT

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PRIMA and is the answer not in the manual?

Questions and answers

New University Dental Centre
May 4, 2025

The time in the autoclave is not synchronized with the current time. How to change the time?

Moayad
January 28, 2025

Error code 7

Subscribe to Our Youtube Channel

Summary of Contents for GALBIATI PRIMA

  • Page 1 QUICK GUIDE FOR ASSISTANT...
  • Page 3 AUTOCLAVE PRIMA ACCESSORIES DESCRIPTION QUANTITY RACK (FOR TRAYS) 1 PIECE TRAYS 4 SET DRAINAGE PIPE 2 PIECES GASKET 1 PIECE INSTRUMENT FOR TRAYS 1 PIECE USER MANUAL 1 PIECE DOOR KEY FOR DOOR REGULATION 1 PIECE USB KEY 1 PIECE...
  • Page 4 AUTOCLAVE PRIMA OVERVIEW A. Function panel B. Clean water tank C. Printer for paper report D. Black door handle E. Blue starting button F. USB port for digital reports...
  • Page 5 FIRST USE OF AUTOCLAVE 1. Open the door strongly pulling the black handle. 2. Take out trays with accessories from the chamber. 3. Remove protective plastic from trays PRIMA and given accessories.
  • Page 6 FUNCTION PANEL A. Shows chamber temperature during a cycle. Unit: °C degrees B. Shows chamber pressure during a cycle. Unit: kPa C. Shows sterilizer period of the cycle, length and warnings. In case of anomalies, the corresponding error code will be shown here. (See User manual, appendix 2, page 42).
  • Page 7 This light is This signal indicates on when the that PRIMA clean water level is a B class autoclave in the tank is low This light is on when This light is on when the...
  • Page 8 II. When it starts, the display will always show the flashing writing «Load» ² and the indicator of class B Prima autoclave will be lightening accompanied by a short sound. Moreover, the indicator corresponding ³ to the last selected program will lighten on the display...
  • Page 9 DISTILLED WATER FILLING NOTE: After switching on the autoclave, if water level in the tank is low the indicator will lighten and an acoustic signal will follow in order to indicate lack of water. So if you want to set up the sterilization cycle you’ll have to add distilled water before doing that.
  • Page 10 DISTILLED WATER FILLING B. ADD BY AUTOMATIC SUCTION: STEP 1: Take the water tube given with PRIMA. STEP 2: › Open the door. › Connect the water tube to the “Distilled water in” connector in the middle by entry connector.
  • Page 11 STEP 2: › Open the door. › Connect the water tube to the “Used water out” entry connector on the left. › Put the other end of the tube in an empty tank or in the available drain system. PRIMA...
  • Page 12 LOADING ARTICLES TO BE STERILIZED ON THE TRAYS 1- Put the articles to be sterilized on the trays so that stream could flow homogeneously. Don’t place the articles on top of each other. (See user manual, page 30) 2- If tools are wrapped it’s better to put them with paper side on the top and don’t overload the trays.
  • Page 13 3 times, temperature: 121°C, pressure: 110kPa, sterilization length 20mins, drying time 18mins. When all the 4 signals brighten the User sterilization program is going to start: standard long cycle at 121°C with length of 40mins more or less. PRIMA...
  • Page 14 SET UP STERILIZATION PROGRAM 1. To set up a sterilization program press “PROG” button. When a sterilization program has been set up by “PROG.” button, its related signal brightens. To change steri- lization program go on pressing 2. Once the sterilization “PROG”...
  • Page 15 In this case, please read the user manual, page 38, appendix 4. 2. Please wait few minutes and open the door. 3. Pull out the trays by using the removal tool in order to avoid any burns. PRIMA INSTRUMENT FOR TRAYS...
  • Page 16 AT THE END OF THE STERILIZATION PROCESS 4. Autoclave provides two methods to save the data of sterilization cycle: A. Through PRINTED REPORT: A the end of the cycle please pull the printed report out from the printer. B. Through DIGITAL REPORT: - Remove the USB from its port and connect the USB in the pc so you can download the digital report.
  • Page 17 TEST PROGRAMS FOR AUTOCLAVE FUNCTIONS HELIX/B&D test lighting signal: VACUUM Test Lighting signal: (see user manual, page 23) (see user manual, page 23) › Press TEST button (1) and › Press TEST button (1) until you select HELIX/B&D test program; select VACUUM test program;...
  • Page 18 POST STERILIZATION CYCLE › Once the sterilization cycle has finished and the working day is at the end, turn off the machine using the switch on/off in the corner down at right on the frontal side of the autoclave. › Make sure to leave the door partially open to let the machine cool.
  • Page 19 B. At least once a week, clean chamber, trays and rack with a clean, damp cloth. C. Clean the mating surface of the chamber with clean, damp cloth. D. Changing door gasket (see user manual, page 33). E. Changing bacteriological filter (see user manual). PRIMA...
  • Page 20 Via Umbria 10 20090 Segrate Milan ITALY info@galbiati.com www.galbiati.com PRIMA...