min., allow drying time of 16 min. Please consider your equipment manufacturer's written
instructions for the specific sterilizer and load configuration being used and current AORN standards
and recommended practices. NOTE: these parameters are for full loads using wrapped sets, rigid
containers and /or peel pouches.
STORAGE INSTRUCTIONS: Store in a cool dry place, and keep away from direct sunlight. Prior
to use, inspect product package for signs of tampering, damage, or water contamination. Use oldest
lots first.
SIGN INSTRUMENTS: SIGN instruments are reusable; however, they have a limited life span.
Prior to and after each use, the instruments must be inspected where applicable for sharpness, wear,
damage, proper cleaning, corrosion and integrity of the connecting mechanisms. Notify SIGN if they
should be replaced. Instrument breakage or damage can occur when an instrument is subjected to
excessive loads, speeds, or dense bone. Striking the cutting surfaces with other metal will cause these
surfaces to become dull.
CLEANING: SIGN instruments and accessories must be thoroughly cleaned before reuse.
Decontamination of reusable instruments should occur immediately after completion of the surgical
procedure. Excess blood or debris should be wiped off to prevent it from drying onto the surface.
Use an enzymatic-cleaning product such as Enzol.
NOTE:
Even surgical instruments manufactured from high-grade stainless steel must be dried
thoroughly to prevent rust formation. All devices must be inspected for cleanliness of
surface and joints, proper function, and wear and tear prior to sterilization.
SHARPENING: The drill bits become dull if they are dinged by hitting the nail or other metal.
They should be protected during surgery, cleaning and sterilization. They are also dulled by pushing
drill bits into bone when they are not advancing. The drill bit heats up and becomes dull.
5
Need help?
Do you have a question about the IM Nail and Interlocking Screw System and is the answer not in the manual?