Sony ODS-L30M Operation Manual page 64

Hide thumbs Also See for ODS-L30M:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 35
Message
I/E STATION
BUSY
I/E STATION
NOT UNLOCKED
I/E STATION HALF-
CLOSED
PUSH I/E STATION
REMOVE CARTRIDGE
FROM CARRIER
ERROR CALIBRATION
REMOVE CARTRIDGE
FROM CARRIER
I/E STATION
OPENED
UNIT CONFIG CHANGE
REQUEST
DIAGNOSTIC LIBRARY
REQUEST
EXPORTED CARTRIDGE
IN I/E STATION
UNIT <unit number>
I/E STATION UNLOCKED
UNIT <unit number>
BULK IMPORT MODE
64
Messages
Description
The I/E STATION button was pressed while the I/E station was being accessed.
Wait for the I/E STATION indicator to light green or orange before pressing the I/E STATION button.
Unlocking of the I/E station failed.
Press the I/E station all the way into the unit so that it is fully inserted, and then try pressing the I/E
STATION button again.
The I/E station is not fully inserted.
Press the I/E station all the way into the unit so that it is fully inserted, and then try pressing the I/E
STATION button again.
The I/E station has been in a state of incomplete insertion for at least 15 seconds.
Press the I/E station all the way into the unit so that it is fully inserted.
Import processing for the I/E station cannot be performed due to the presence of a cartridge in the
carrier.
Move the cartridge from the carrier to an empty slot via the [Move Cartridge] tab of the Library
Maintenance Web UI.
Carrier adjustment failed due to the presence of a cartridge in the carrier. Consult your local Sony
representative.
An I/E slot movement prompt was issued while the I/E station was open.
Close the I/E station, or check the application controls.
A change in the unit configuration was detected.
Consult your local Sony representative.
Library diagnostics are required.
Consult your local Sony representative.
A cartridge was exported to the I/E station of the displayed unit. Remove the cartridge from the I/E
station as required.
The I/E station of the displayed unit is unlocked. Open the I/E station, exchange the cartridge, and
then close the I/E station.
[Bulk Import] is set to "ON" on the [Import/Export Cartridge] tab of the Library Maintenance Web UI.
After cartridge import processing is finished, always set [Bulk Import] to "OFF."