Safety Precautions The Blood Bank unit is designed for CAUTION storage of whole blood and related products at constant temperature in clinical, pharmaceutical, research, and This action may cause fatal injury other laboratory applications. Please or servere damage. carefully read this manual, and follow the instructions to properly operate the unit.
Page 4
SAFETY PRECAUTIONS WARNING Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain water. Only qualified engineers or service personnel should install the unit. The installation by unqualified personnel may cause electric shock or fire. Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from turning over.
Page 5
SAFETY PRECAUTIONS WARNING Use this unit in safe area when treating the poison, harmful or radiate articles. Improper use may cause bad effect on your health or environment. Turn off the power switch (if provided) and disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent electric shock or injury.
Page 6
SAFETY PRECAUTIONS CAUTION If the unit is to be stored unused in an unsupervised area for an extended period, ensure that children do not have access and that doors cannot be closed completely. The disposal of the unit should be accomplished by appropriate personnel. Remove doors to prevent accidents such as suffocation.
Installing the appliance Always allow an authorised electrician to install the appliance. The appliance should be located in a cool, dry place out of direct sunlight. The appliance gives off a great deal of heat when in operation. If the ambient temperature is too high there is a risk that the appliance will not operate correctly.
Descriptions of Refrigerator Parts and their Functions 3. Battery Back-up 7. Chart Recorder 11. Rating plate 1. Control panel 9. Interior Light 4. Door lock with key 2. Light Power Switch 5. Door Handle 8. Blood Storage Drawers 8. Adjustable Devices 6.
Page 9
Descriptions of Refrigerator Parts and their Functions 1. Control Panel ( lle ) The control panel on the blood bank shows the actual storage temperature inside the refrigerator. You can use the push buttons on the panel to adjust the temperature set point and test the alarm functions. 2.
How to mount the Door Handle 1. Door Handle with Fittings and Cover 2. Place the Fittings on the Door Handle. 3. Mount the Handle on the door with screws. 4. Push the cover for the Door Handle on its place. 5.
Location of Refrigerator Your refrigerator should be placed on a location that satisfies the following conditions. By doing so, the optimum operating results can be achieved. 1. Firm, level floor surface Installing the unit on a firm and flat floor surface WARNING reduces the chance of excessive noise and vibration.
Installation Remove packaging materials, packaging bags and straps. Open the refrigerator door to let it air out. If there are dirt marks on the outside of the unit, use a soft sponge with a light detergent solution to wipe the unit clean. Then, rinse the unit with a clean, slightly damp towel.
Up, down, left, right button (Dixell XR30CX) to scrolls through the menu On/off or exit menu Enter (accesses the menu, accept) Mute alarm Alarm (red light) Description control panel for user Screen At alarm Main menu Current alarm Non current alarm(s) Temp.
Description control panel for engineer Screen Main menu At alarm Temp. Last 24H max/min Settings level l password Settings level ll password ll Hysteresis Range Max temp limit 13 Hysteresis Min temp limit The difference between the set point and the temperature that causes the fan and the compressor to restart.
Operation 1. Inspect to see if the unit is overloaded or products with warm temperature have been loaded into the unit. 2. Make sure there is an air gap between products. Unit’s refrigeration is not effective. 3. Check to make sure the unit is not in direct Temperature tends to operate out of range.
Chart Recorder 1. Diagram 2. Pointer 6. Fibre point 7. Battery 9. Set of keys 10. Set of spare screws 3. Door 4. Diagram holding device 5. Door lock 8. 100 diagrams The Chart Recorder is device intended to monitor the temperature and/or the pressure. It is made up of a neutral gas thermometer and write on a circular diagram with a fibre point pen.
Page 20
To achieve the operation: Open the recorder door (3). Lift the pointer (2) so as to release the diagram (1). Unscrew the knurled nut (4) and remove the diagram (1). Firmly hold the case in one hand and, with the other, pull the mechanism axis toward the outside by alternatively swinging up and down to release it from its base.
Page 21
After rewinding, lift the pointer (2). Unscrew the winding knob and the diagram´s locking knob (4). Take a blank diagram (8), note the date and identify the recording in the boxes foreseen so that effect. Install the diagram on the axis and slip the edge of the disk under the 2 holding tabs. Screw back knob, clockwise, up to the stop and then come slightly backwards so as to be able to make the diagram turn.
Page 22
When you are running short of the chart paper, please contact Arctiko to purchase more. The lead time to deliver the charts is 1-3 working days after receiving the order. Please call Arctiko for details.
Proper usage of the Refrigerator During the initial startup and continuous usage of the refrigerator, the following procedures should be followed. 1. Power the refrigerator in a dedicated power socket. The voltage requirement is 220 VAC and 50 Hz. When started, do not place any products in the refrigerator. 2.
Maintenance WARNING Before any inspection or maintenance work is performed, the refrigerator’s power plug should be disconnected from the power supply socket. This is to prevent any potential electrical shock or injury. During the maintenance work, do not breathe the dust and aerosols near the unit; they might be harmful to your health.
How to change the flourescent light tube Unplug the refrigerator from the power supply. Remove the cover for the fluorescent light tube. Turn the tube and pull it out the split. Push and turn the new tube in its place. Install the cover again.
Specifications Description Blood Bank Refrigerator For Medical Application Model Number BBR100 BBR300 BBR500 BBR700 BBR1400 W 610 x W 520 x W 620 x W 720 x W 1440 x D 645 x D 700 x D 860 x D 860 x...
Page 27
70 kg 178 kg Rechargable battery DM8-6 (8V6AH) Operating Parameters Model BBR100 BBR300 BBR500 BBR700 BBR1400 Sensor Bottle Temperature +4° C +/- 2° C Design Ambient Temperature +5° C To 35° C Sustainable Alarm Time 72 Hours With Fully chargeable battery...
Rating Plate The Rating Plate is placed inside the machinery room on the right side. Logo Address and tlf.no. Modelname Compressor type El. current (A) Voltage Symbol for Rosh. CE sign for Declaration and Conformity Made in EU 10. EN standard 11.
10 years from purchase date. Guarantee ARCTIKO provides a 100% guarantee on parts for a period of 12 months, excluding shipping costs and labor. There is no guarantee for the contents of the unit.