Summary of Contents for IKA CBC 5 control

  • Page 1 20000013222 CBC 5 basic_112016 CBC 5 control ® Operating instructions...
  • Page 2: Device Setup

    Device setup RS 232 Micro USB Pt 100 Multi I/O Fig. 1 Item Designation Wireless Controller (WiCo) Filling opening lid Front cover Handle Venting grid Mains switch Power socket Cable clip Pump connection IN Pump connection OUT RS 232 interface USB interface External temperature sensor socket Multifunction interface...
  • Page 3: Table Of Contents

    Connecting the tubings ���������������������������������������������������������������������������������������������������������������������������������������������8 Inserting battery into the WiCo ����������������������������������������������������������������������������������������������������������������������������������9 Mounting the WiCo to the station �����������������������������������������������������������������������������������������������������������������������������9 Filling and draining ����������������������������������������������������������������������������������������������������������������������������������������������������9 Fluid (Standard information for IKA fluid)���������������������������������������������������������������������������������������������������������������10 ® Moving the device ���������������������������������������������������������������������������������������������������������������������������������������������������11 Charging the Battery Pack RB 1 (rechargeable battery) ���������������������������������������������������������������������������������������������11 Changing battery to WiCo ���������������������������������������������������������������������������������������������������������������������������������������11...
  • Page 4: Declaration Of Conformity

    Declaration of Conformity We declare under our sole responsibility that this product corresponds to the regulations 2014/35/EU, 2014/30/EU, 2014/53/EU and 2011/65/EU and conforms with the standards or standardized documents: EN 61010-1, EN 61010-2-010, EN 61010-2-051, EN 61326-1 and DIN 12876-1� Bluetooth module: ®...
  • Page 5: Safety Instructions

    70 ° C� plosions or comprehensive monitoring equipment� • Process pathogenic material only in closed vessels under a It is dangerous to touch the heater� suitable fume hood� Please contact IKA application sup- ® CAUTION The temperature of the heater can port if you have any question�...
  • Page 6: Fluids

    • After a power failure during operation, the device may Risk of burning caused by vapor or CAUTION start automatically (depending on operating mode)� hot water at the outlet of the cool- • Transport the device with care� ing coil� •...
  • Page 7: Correct Use

    • Explosion of a battery pack may release battery fluid and Disposal instructions: cause a fire� • When disposing of the IKA battery pack, please tape over ® the contacts with adhesive tape to prevent short-circuiting • The lithium polymer battery pack must only be used and due to moisture or contact with metal components�...
  • Page 8: Unpacking

    Open the filling open lid (2) and remove the protection under the Fig. 2 Fig. 3 buoyage� • Delivery scope: - CBC 5 control station with WiCo - Mains cables - Hose olive NW 12 (2 pieces) see Fig� 2 Adapter Adapter - Screwdriver (use for safety circuit) see Fig�...
  • Page 9: Inserting Battery Into The Wico

    • Inserting battery into the WiCo: - Check and make sure that the drain valve is closed (Ro- tate clockwise to the stop position, see Fig� 10)� Note: Please note information in section “Commissioning”� - Connect the mains plug and turn on the device with mains switch (6)�...
  • Page 10: Fluid (Standard Information For Ika ® Fluid)

    • Fluid (Standard information for IKA fluid): ® ® Operating temperature Operating temperature Safety temperature Flash point Designation range for open bath range for closed bath (°C) (°C) application applications (°C) (°C) CF�EG28�N10�80�8 -10 ��� 80 -10 … 80 CF�EG39�N20�80�16 -20 …...
  • Page 11: Moving The Device

    • Moving the device: Empty all fluid in the bath before moving device from one place to other place� The device must be lifted by two persons with the upper handles� For this purpose, take off the front cover (3)� It can also be moved on flat surface by lifting and pushing the front of the device�...
  • Page 12: Operator Panel And Display

    Operator panel and display • station: Fig. 15 Item Designation Function On/Off key: Switch on/off the station� Adjustable safety circuit: Adjust the safety temperature limit with delivered screwdriver� LED bar: Display different status of the circulator with different color� Contacts: Communicate and charge the WiCo�...
  • Page 13: Setting The Safety Temperature

    Setting the safety temperature Seting the safety temperature with screwdriver delivered The safety temperature setting will appear on the display� with the device� Factory setting: maximum value Adjustment range: see “Technical data“� Note: The safe temperature limit must always be set to at least 25 ºC lower than the flash point of the fluid used�...
  • Page 14: Useful Information

    Note: In standby status, press the right knob (R) to start the Note: In an emergency, the device function can be turned pump function, the heating/refrigerating function will not off by pressing the “safe STOP” key (E) at the front of sta- be activated�...
  • Page 15: Working With Wico

    Working with WiCo • Working screen at the time of delivery: Operating Mode: This symbol indicates the operating mode currently select- ed (A, B, C, D)� 0�00 USB: This symbol means the WiCo is communicating or the int °C battery is charged via a USB cable� The symbol no longer appears if no USB cable is being used for communicating with the station�...
  • Page 16: Menu Navigation And Structure

    • Menu navigation and structure: Menu navigation: Press the "Menu" key (U)�  Select the menu by turning the right rotating/pressing knob (R) to the right or  left� Open the menu item by pressing the right rotating/pressing knob (R)� ...
  • Page 17 Menu structure: Factory Settings TEMPERING Control Control Mode Internal (int) activated External (ext) Control Parameters Automatic Accurate activated Fast Internal (Kp, Ti, Td, Ts, Prop_Bp, Prop_Bn) Manual 60.0, 5.0, 0.3, 3, +1.00, -1.00 External (Kp, Ti, Td, Ts, Prop_Bp, Prop_Bn) 1.0, 15.0, 0.0, 90, +1.00, -1.00 Information (Kp, Ti, Td, Ts, Prop_Bp, Prop_Bn) Fluids...
  • Page 18: Menu (Details)

    The differential time Td (s) is the derivative time and de- temperature� See table in the section “Fluid (Standard in- termines how strongly the rate of change of the control formation for IKA fluid)“� ® deviation affects the control variable� Td compensates for The maximum and minimum temperature values of the rapid control deviations�...
  • Page 19 Limits: Under the option "External (ext)", the maximum and min- Under “Power Output“, the maximum heating output imum temperature for external temperature control can can be set as a percentage of nominal heat output� be limited� Under "l Δ T (int - ext) l", the maximum difference be- tween the internal and external temperature can be set�...
  • Page 20 Manual: TIMER Axis Assignment: Set: The temperature values to be displayed can be selected� You can set a target time (duration) in this menu option� Axis scaling: When device functions are started normally, this time is Scaling of the time (X) and temperature axis (Y) can be displayed on the working screen�...
  • Page 21 M2 (ON/OFF): MODULES M2-Output switch Note: Prior to using the “Refill” menu option, check the OFF: M2-output switch in initial state� function of the buoyage� ON: M2-output switch in inverted initial state� The “Refill” menu option has higher priority than the pro- gram M1 segment settings�...
  • Page 22 M3-Input: SETTING Ext. Standby: Languages: External standby input to stop the device functions Tem- The “Language“ menu option allows you to select the per and Pump� desired language� ON: Activate the ext� standby function� Device functions are stopped in the event of an inverted initial state (ON) Display: at the input�...
  • Page 23: Interface And Output

    1�1)� The NAMUR commands and the additional specific IKA co mmands serve only as low level commands for communication ®...
  • Page 24 IN_TMODE Read temperature control 0: internal control 1: external control OUT_SP_1 xxx Set the internal setting temperature XXX (if 0: internal control) Set the external setting temperature XXX (if 1: external control) OUT_SP_12@n Set the WD safety temperature with echo of the set (defined) value� OUT_SP_4 xxx Set the pump speed XXX OUT_SP_42@n...
  • Page 25 Connection WiCo to station: USB micro B USB micro A USB micro B USB micro A USB cable Fig. 29 Connection the device to PC: 9-pin RS 232 9-pin RS 232 RS 232 cable USB cable USB micro B USB A USB micro B USB A USB micro B...
  • Page 26: Maintenance And Cleaning

    - serial number, see type plate • Clean condenser dust filter with a vacuum cleaner or - item and designation of the spare part, see www�ika�com wash the filter in the water and dry it before assembly� - Software version�...
  • Page 27: Error Codes

    Error codes Any malfunctions during operation will be identified by an error message on the display� Proceed as follows in such cases:  Switch off device using the main switch at the back of the device  Carry out corrective measures ...
  • Page 28: Accessories

    USB 2�0 cable micro A - micro B H�FKM�8 FKM tube (nominal width 8 mm) USB 2�0 cabel A - micro B H�FKM�12 FKM tube (nominal width 12 mm) Labworldsoft ® Tubing insulations: ISO� 8 Insulation (8 mm) ISO�12 Insulation (12 mm) See more accessories on www.ika.com�...
  • Page 29: Technical Data

    Technical data station Operating voltage 230 ± 10 % / 100–115 ± 10 % Frequency 50 / 60 Max� input power 3000 (230 V) / 1750 (115 V) Working temperature range °C - 25 ��� + 200 Operating temperature range °C - 25 ���...
  • Page 30 WiCo Permitted on time Max� communication distance (dependent on the building) Dimensions (W x D x H) 160 x 40 x 105 Weight 0�3 Ambient temperature + 5 ��� + 40 Ambient humidity (relative) Protection class according to EN 60529 IP 40 Interface RB 1 Battery pack...
  • Page 31: Warranty

    Warranty In accordance with IKA warranty conditions, the warranty The warranty does not cover worn out parts, nor does it ® period is 24 months� For claims under the warranty please apply to faults resulting from improper use, insufficient care contact your local dealer�...
  • Page 32 -Werke GmbH & Co�KG ® Janke & Kunkel-Str� 10 D-79219 Staufen Tel� +49 7633 831-0 Fax +49 7633 831-98 sales@ika�de www�ika�com...