Setup The Ebb technology is based on extensive clinical data and your sleep by trying Ebb. 20+ years of sleep imaging research. Connect Headband Insert Fluid Cartridge Relief from sleeplessness can take time and commitment.
– If the enclosure of the bedside controller is broken. doctor before use. • There are no user-serviceable components inside the device. Do not • Do not use Ebb if the package or any part is damaged. Contact Ebb attempt to open the bedside controller. Customer Care.
Cautions alert the user to important information regarding use of Ebb. • Improper placement of the headband can result in reduced therapy. • The safety and effectiveness of Ebb in pregnant women has not been established. • Use only fluid cartridges provided by Ebb Therapeutics to assure proper performance.
Unpack and familiarize yourself with Ebb. Ebb is the first smart-technology and drug-free wearable device that uses a proprietary cooling headband to naturally trigger a reduction in brain activity, which quiets the racing mind to help people with sleeplessness to fall asleep faster and get in to deeper stages of more restorative sleep faster.
Keep the power supply away from hot surfaces. • Ebb is isolated from the mains by means of the supplied medical grade power supply. Do not use a power supply other than the one supplied with the system.
Wait 30 seconds before using bedside controller. CAUTION: Use only fluid cartridges provided by Ebb Therapeutics to assure proper performance. Make sure the fluid cartridge is in the bedside controller WARNING: before initial use. Do not use Ebb if the package or any part is damaged. Contact Ebb Customer Care Team.
Display Set the Clock When in use at night, the display disappears. Rotate the dial to wake the display. When the clock is displayed, rotate counter-clockwise Clock/Alarm Buttons on/off on/off to dim the display and clockwise to make clock alarm gentlewake it brighter.
Page 9
Set Wake Time Turn on Gentlewake 7:00 Press “Gentlewake on/off” button. A sun The hours begin blinking. Turn 8:30 icon will appear when the Gentlewake Open lid to access clock buttons. Press and release the “Set Alarm” button. dial to set hour with AM or PM. Open lid to access clock buttons.
• Some individuals may experience skin redness on the forehead after wearing the forehead pad. If this persists more than 12 hours, contact Ebb Customer Care. • The headband must be worn in the proper orientation to prevent entanglement and strangulation.
Ebb. WARNING: Do not drop or place Ebb in any fluids. Do not allow any fluids to enter the openings of Ebb. Precaution: If you notice a change in functioning of Ebb, please contact Ebb Customer Care.
The expected service life of the Ebb system is 3 years. If the device Separate collection for electrical and electronic equipment per The circulating fluid in Ebb naturally evaporates. After about three months of use, the fluid stops cooling or circulating the fluid or begins to leak, it is time to EC Directive 2002/96/EC.
Check for kinks or bends in the tubing. There is a problem with the bedside controller Disconnect the power cord and then plug back in. If error code repeats, contact Ebb Tubing disconnected Ensure tubing fully connected into back of bedside controller.
Degree of Protection Against Electric Shock: Type BF Applied Part Degree of Protection against Ingress of Water: Keep Dry Isolation from Supply Mains Ebb is isolated from the mains by means of the supplied medical grade power supply. Mode of Operation: Continuous Rated Power...
Page 15
Ebb complies with all applicable electromagnetic compatibility requirements (EMC) IEC 60601-1-2 4th edition and meets the Class B Group 1 Ebb is intended for use in the electromagnetic environment specified below. The customer or the user of Ebb should assure that it is used emissions requirements in CISPR 11.
Page 16
Guidance and manufacturer’s declaration – electromagnetic immunity Ebb is intended for use in the electromagnetic environment specified below. The customer or the user of Ebb should assure that it is used in Ebb is intended for use in the electromagnetic environment specified below. The customer or the user of Ebb should assure that it is used in such an environment.
Page 17
Recommended separation distances between portable and mobile RF communications equipment as well as RF wireless communications equipment Ebb Ebb is intended for use in the electromagnetic environment specified below. The customer or the user of Ebb should assure that it is used in such an environment.
Class B digital device, pursuant to part 15 of the FCC Rules. These If the Product fails under normal conditions of use and the Product is returned to Ebb Therapeutics within the Warranty Period, Ebb Therapeutics limits are designed to provide reasonable protection against harmful will repair or replace the defective Product with a new or refurbished Product, at Ebb Therapeutics’s sole discretion.
Need help?
Do you have a question about the Ebb and is the answer not in the manual?
Questions and answers