Page 1
SERIES Digital Auto Sensor Alternating Air- Pressure Mattress System USER MANUAL DX550 Grade MANUAL DX550 Alternating Air-Pressure Mattress Replacement Syste m (Digital) Grade * This ‘Grade’ is a guide only. Each individual must be assessed by a tissue viability specialist, the correct product chosen and set to operate according to the patient’s needs.
1. The Purpose of this Manual 1. The Purpose of this Manual This operation manual is mainly focused on the setup, cleaning and routine maintenance of the Steiss This operation manual is mainly focused on the setup, cleaning and routine maintenance of the Steiss DX550 Alternating Air-Pressure Mattress Replacement System.
• Use only manufacturer-approved parts when maintaining or repairing the Steiss DX550 • Use only manufacturer-approved accessories with this device. All parts and accessories supplied are specifically designed for use with the Steiss DX550 control • unit. • Use of other parts and accessories in conjunction with this device is not recommended.
This will remind the care team when the annual service is due. 3. Hang the Steiss DX550 pump at the foot of the bed. Connect the push-fit hose connector from the foot of the mattress to the air outlet of the pump.
INFLATION OF MATTRESS 1. Ensure that the Cardio Pulmonary Resuscitation (CPR) valve of side mattress is closed. CAUTION: It is important to ensure that no person/object is lying/sitting on the mattress when it is initially being maximum pressure. 4. When the mattress is fully inflated, the mattress is now ready to be used and patient can be placed on to the air mattress.
MATTRESS CPR VALVE CHECK After approximately 30-40 minutes the mattress should be fully inflated. If the mattress is NOT fully inflated, check that the CPR valve is in the closed position and that the pump is connected to the mattress correctly. "...
4. Emergency Air Release – 4. Emergency Air Release CPR Valve Function - CPR Valve Function This mattress features a Cardio Pulmonary Resuscitation (CPR) valve at the head-end side of the mattress. In the event of a cardiac arrest, twist the CPR valve to the ‘open’ position and unplug the air hose from the IMPORTANT: ALARM FUNCTION The DX550 Alternating Air - Mattress Replacement System is equipped with visual and audible alarms in the...
For added durability, and longer working life, your Steiss™ DXR500 or added durability, and longer working life, your Steiss™ DXR500 For added durability, and longer working life, your Steiss™ DXR500 For added durability, and longer working life, your Steiss™...
• • Use a Steiss 5 pump with a Steiss M 5 0 mattress only. Use a Steiss 5 pump with a Steiss M 5 0 mattress only. • • When cleaning do not use Phenol based substances. Steiss offer Multi-surface disinfection wipes When cleaning do not use Phenol based substances.
Check air cells for holes/tears (other Replace any damaged air cells. than where part of the design) Consult a qualified service technician if the problem persists. If issue not resolved, please contact your Steiss Supplier.
Mattress cells can be wiped down with a solution of sodium hypochlorite 1,000ppm or any other non-phenol- based germicidal cleaning solution. Steiss have developed a dedicated wipe to ease the cleaning process for their clients. Please ask your mattress supplier for Multi-surface disinfection wipes.
The pump casing is manufactured from ABS plastic - therefore: NOTE: Do not use a phenol-based cleaning solution. Steiss has designed cleaning wipes called Multi- NOTE: Do not use a phenol-based cleaning solution. Steiss has designed cleaning wipes called Multi- surface disinfection wipes.
Page 13
10. Storage and Care of all parts 10. Storage and Care of all parts MATTRESS • Check the air manifold for kinks or breaks and replace it if necessary. Turn the CPR valve to the “OPEN” position and disconnect the air tubes from the control unit to allow the •...
11. Specification/Technical Data Steiss 5 Master Control Unit / Pump Model Name Steiss 5 Master Control Unit Size L x W x H 34.5 x 14.5 x 21cm (13.58 x 5.71 x 8.26 inches) Weight 3.2kg (7.05lbs) Cycle Time 10 minutes...
Page 15
Recommended separation distances between portable and mobile RF communications equipment and the Steiss DX550 The Steiss DX550 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Steiss DX550 can help prevent electromagnetic interference by...
Page 16
Guidance and manufacturer’s declaration – electromagnetic emissions The Steiss DX550 is intended for use in the electromagnetic environment specified below the customer or the user of the Steiss DX550 should ensure that it is used in such an environment. Emissions Test Compliance Electromagnetic environment –...
Page 17
Guidance and manufacturer’s declaration – electromagnetic immunity The Steiss DX550 is intended for use in the electromagnetic environment specified below. The customer or the user of the Steiss DX550 should ensure that it is used in such an environment. Immunity test...
Page 18
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Steiss DX550 is used exceeds the applicable RF compliance level above, the Steiss DX550 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the product.
15. Expected Service Life 15. Expected Service Life The Steiss DX550 pump has an expected minimum service life of three years. The Steiss DX550 pump has an expected minimum service life of three years. The Steiss DX550 pump has an expected minimum service life of three years.
These Descriptive Risk Levels should not be taken as prescriptive criteria. C. Steiss mattresses/cushions should be seen as an aid to care and DO NOT replace the need for good C. Steiss mattresses/cushions should be seen as an aid to care and DO NOT replace the need for good nursing care and intervention.
18. Service 18. Service RECORD PRODUCT DETAILS Serial Number Model Number Ward or Care Establishment Deployment date / first opening date SERVICE EQUIPMENT Flow gauge Manometer Details of mattress and pump audit procedure SERVICE GUIDELINES General Condition - check these areas: Details: PU Cover, Cover Welds, Screen Print and Base Substrate Look for signs of damage...
NOTE: This is only a physical check of the mattress / pump system and by no way meant to replace anything that is recommended by the MHRA EPUAP or similar professional body. We recommend you check your relevant Healthcare Body website for any further information that may need to be incorporated into your checks.
Page 23
Service Record Service Record First service due: First service due: Steiss Authorised Service Centre: Steiss Authorised Service Centre: All applicable service actions checked: All applicable service actions checked: Open service actions completed: Open service actions completed: Date: Date: Genuine Steiss** parts used:...
Page 24
Supplier Contact Details Last Edited June 2018...
Need help?
Do you have a question about the 5 Series and is the answer not in the manual?
Questions and answers