In-situ SMARTROLL MP Instruction Sheet

Storage and calibration cup
Hide thumbs Also See for SMARTROLL MP:

Advertisement

Quick Links

TROLL™ MP Storage and Calibration Cup
SMAR
Part Number 0093232
Application
The
TROLL MP Storage and Calibration cup is a
SMAR
rugged alternative to the standard calibration cup. The
calibration procedure for this cup is similar to the
instructions in the Operator's Manual with a slight
variation in the way the cup is prepared prior to
calibration.
Kit Contents
Polycarbonate calibration cup
l
Sponge wafers
l
Information subject to change without notice. In-Situ, In-Situ logo, BaroMerge, BaroTROLL, HERMIT, iSitu, Pocket-Situ, RDO,
RuggedCable, RuggedReader, TROLL, and Win-Situ are trademarks or registered trademarks of In-Situ Inc.
Instruction Sheet
iSitu App
1. Calibrate the sensors using the iSitu App running on
an iOS device.
2. Tap the Calibration icon
access a list of sensors that are available for
calibration.
RDO 100 Percent Oxygen Saturation
Calibration
The RDO sensor requires very little maintenance. The
zero oxygen calibration is optional. Complete this
calibration before you calibrate the remaining sensors.
1. To perform the 100 percent oxygen saturation
calibration, remove the bottom of the calibration cup
and place a water saturated sponge in the end cap.
2. Attach the end cap and sponge to the cylinder.
©
2013. All rights reserved.
in the iSitu App to
04/2013 | Rev. 001 | 0093232

Advertisement

loading

Summary of Contents for In-situ SMARTROLL MP

  • Page 1 2. Attach the end cap and sponge to the cylinder. Information subject to change without notice. In-Situ, In-Situ logo, BaroMerge, BaroTROLL, HERMIT, iSitu, Pocket-Situ, RDO, 04/2013 | Rev. 001 | 0093232 ©...
  • Page 2 3. Set the instrument in the calibration cup. 4. Follow the calibration directions in the Operator's Manual or in the Mobile Help to complete the procedure. Do not fasten the instrument to the cylinder. The chamber must be vented to the Instruction Sheet atmosphere to produce a valid calibration.