Installing actuator
1. Reinsert fixing bolts and SL retainer clips. Fix SL
retainer clips securely to fixing bolt with the hammer.
2. Secure actuator cable to the frame with cable ties.
The cable must be routed so that it does not rub or
get trapped.
3. To complete, check all seat tilting functions.

5.13.3 Replacing Seat Height Adapter

The adapter raises the seat-to-floor height of a Modulite
seat. This allows you to mount the seat further to the
back of the mobility device.
Installation kit
13 mm wrench (2x)
1. Remove legrests.
2.
Release seat A and tilt forwards.
3. Separate cables from power module.
1528338-Q
Fig. 5-77
4.
Loosen and remove screws B.
5. Remove seat frame support C.
6.
Use long screws, delivered with installation kit,
to fasten seat frame support.
Install parts in reverse order.
7. Test all functions.
Fig. 5-78