Keithley 3933A Operator's Manual page 54

Phase shifter
Table of Contents

Advertisement

APPENDIX
I3
Model 3933A Specifications
B.l ELECTRICAL SPECIFICATIONS
I Waveforms
Oscillation
Modes
Set by 3930A mode.
Frequency
Set by 3930A frequency.
Waveform
and Frequency Range
2,
, n
(Duty cyde fixed at 50%)
$ ,/l
,U
, 1
(Duty cycle varies
from 5% to 95%)
O.lmHz to 1.2MHz
0.1mHz to lOOkH.2
Phase
Setting Range
Display
Accuracy (when 3930A is in
CONT mode)
-360" to 360" (Value corresponding
to 0" at 3930A's start/stop
phase)
Maximum
4 digits f resolution
0.1' (futed)
o Set both 3930A and 3933A to these settings:
DC offset OV, AM off, 5OQ load, same waveform,
18°-280C.
o Waveform
duty cycle is fixed or variable on both devices. (cycle is optional)
o Amplitude
setting is between 3O.OmVp-p to 3O.OVp-p,
and can be set independently.
o When connecting in sequence, phase number n is per this diagram:
3930Al
3933A [
13933Al
. . . . . .
n=l
n=2
n=3
However,
n < 6 (to max of 6 phases)
'
Accuracy corresponds
to:
(3933A phase setting) - (3930A's start/stop
phase setting)
D = + 0, -(n-2) x 40ns
f-b
ffO.1" + 12011.5) + D
rt
When duty cycle is 50% (FXD)
zt(O.5" + 12011s) + D
C
KO.1" + 22011~) + D
(while rising)
%
i(O.1" + 9Ons) + D
n
(while falling)
ic(O.1' + 24011s) + D
\
(while rising)
Specifications
subject to change without
notice.