350VL Patient Quick Reference
This guide is meant as a complement to the Operation Manual.
Please refer to the Operation Manual for complete instructions
for use.

Display

Sounds / Lights

Message

Sounds: Slow beeping
Lights:
Blinking yellow
Sounds: Slow beeping
Lights:
Blinking yellow
Sounds: Continuous tone
Lights:
Continuous red
Sounds: Continuous tone
Lights:
Blinking red
Sounds: Continuous tone
Lights:
Blinking red

Condition

Not delivering

Low battery

Depleted battery

Clamp bar is open

IV tubing occlusion

Continued on following page

Corrective Action

Press
key to start delivery.
Press
key to silence alert for 5 minutes.
Press
key to start delivery.
Replace battery within 30 minutes.
Remove depleted battery to silence alarm.
Replace battery.
Press
key to turh pump on.
Press
key to restart.
Press
key to silence alert.
Close clamp bar.
Press
key to restart.
Check tubing system for kinks and correct.
Pump will resume delivery when line is cleared.
Press
key to silence alert.
Check tubing system for kinks and correct.
Press
key to restart.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 350VL and is the answer not in the manual?

Questions and answers

Summary of Contents for WalkMed 350VL

  • Page 1 350VL Patient Quick Reference This guide is meant as a complement to the Operation Manual. Please refer to the Operation Manual for complete instructions for use. Display Sounds / Lights Condition Corrective Action Message Sounds: Slow beeping Not delivering Press key to start delivery.
  • Page 2 - or - Set total volume limit to OFF. Failed key in keypad Press key to turn pump off. Take pump out of service. Sounds: Continuous tone System malfunction during Remove battery and wait 15 seconds before re- battery install installing. Lights: Continuous red © 2018 WalkMed www.walkmed.com/resources/wm-350vl-patient-quick-reference/...

Table of Contents