Advertisement

Quick Links

Tinytag Transit 2
Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Transit 2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Tinytag Transit 2

  • Page 1 Tinytag Transit 2 Manual...
  • Page 3: Table Of Contents

    C O N T E N T S Warnings ............4 Getting Started .
  • Page 4: Warnings

    W A R N I N G S If this equipment is installed or used in a manner not specified by the manufacturer • then the protection provided by the equipment may be impaired. This equipment contains a lithium battery. Danger of explosion if the battery is •...
  • Page 5: Getting Started

    You will first need to install the Tinytag Explorer software and the USB inductive pad or cable as described in the Tinytag Explorer quick start guide. To use the Tinytag Transit 2 with an inductive pad it is simply as case of positioning the logger in the centre of the pad as shown below.
  • Page 6: Features

    F E A T U R E S Memory Type The Tinytag Transit 2 has a non-volatile memory, which means that the unit’s battery can be replaced, once the unit has been stopped, and the data held on it will be retained.
  • Page 7 Trigger Start The Tinytag Transit 2 has a trigger start option that allows a user to set up a data logger as they wish it to record, and then start it remotely at a later time by passing a magnet across the back of the unit.
  • Page 8 F E A T U R E S Delayed Start Setting a delay start gives the user time to get a unit to the measurement location before it starts logging. Two type of delay can be set, a relative time (e.g. one hour from now) or an absolute time (e.g.
  • Page 9: Led Flash Patterns

    F E A T U R E S Alarms The TG-4080 has a red LED that can be programmed to flash when the temperature exceeds limits set by the customer. The user can set two independent values to trigger the flashing of the red LED, and both values can be set to latch so that a unit will continue to show an alarm even when the temperature goes back to acceptable values once again.
  • Page 10: Specification

    S P E C I F I C A T I O N Reading Range -40° C to +70° C (-40° F to +158° F) Reading Accuracy <0.4° C between 0 and 50° C Reading Resolution 0.01° C or better Response Time 10 mins to 90% FSD in moving air IP Rating...
  • Page 11: Calibration

    We recommend that the calibration of this unit should be checked annually against a calibrated reference meter. A UKAS traceable certificate of calibration can be supplied if the unit is returned to Gemini for a service calibration. For further information see: www.tinytag.info/calibration...
  • Page 12: Battery Replacement

    If used at low temperatures the data logger should be allowed to warm to room temperature before it is opened to avoid condensation forming inside the unit. A service kit for the Tinytag Transit 2, containing a replacement battery and servicing instructions, can be ordered from your supplier using the...
  • Page 13: Approvals

    A P P R O V A L S This logger complies to BS EN 12830, between -30 and +30° C, in the following categories: S; T; C; D; 1 This equipment complies with part 15 of the FCC rules. Operation is subject to the following two conditions: (1) this device may not cause any harmful interference, and (2) the device must accept any interference received, including interference that may cause undesired operation.
  • Page 14: Warranty

    W A R R A N T Y This product carries a manufacturing defects warranty of 12 months from the date of • purchase. Units returned under warranty will be repaired or replaced at the manufacturer’s discretion. This warranty does not cover mishandling, modification or battery replacement and is subject to the standard Terms and Conditions of Sale, a copy of which is available on request.
  • Page 15: Further Information

    F U R T H E R I N F O R M A T I O N Further information on Tinytag data loggers, software and accessories can be found on our web site at: www.tinytag.info If you should have any further questions, please contact your distributor or: Gemini Technical Support t: +44 (0)1243 813009 e: help@tinytag.info...
  • Page 16 Gemini Data Loggers (UK) Ltd. Scientific House, Terminus Road, Chichester, West Sussex, PO19 8UJ England. www.tinytag.info t: +44 (0)1243 813000 e: sales@tinytag.info 9800-0040: Issue 6 (18th July 2014)