TABLE 2:
GUIDANCE AND MANUFACTURER'S DECLARATION — ELECTROMAGNETIC IMMUNITY
InTENSity at Home
electrical stimulators are intended for use in the electromagnetic environment specified below.
™
The customer or the user of these electrical stimulators should assure that it is used in such environment.
Immunity test
Electrostatic
discharge (ESD)
lEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC
61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
IEC 60601 test level
±6 kV contact
±8 kV air
Not applicable
Not applicable
Not applicable
3 A/m
Compliance level
Electromagnetic environment - guidance
±6 kV contact
Floors should be wood, concrete or ceramic tile.
±8 kV air
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
39