Z2000 / C2020
To Calibrate the Gauge, Open the front cover 4 screws, Connect the Tank
1
Probe, and Power the unit.
Toggle the calibration switch, top right marked CAL (LK2)
The first screen to show is a "Cube Tank"
2
Hidden Button
Hidden Button Scrolls through settings.
4
You Must Scroll to YES and use the
button to
Confirm YES for all Screens
You must scroll to YES and
Confirm YES with Left
6
hand Hidden Button to
save your values.
Z2000/C2020 Tank Gauge Setup Guide – PD02/0001 Rev.01 Feb/2017
Tank Gauge Set-Up
Middle Button Scrolls through Tank Type
3
Choose the Feature you require, then set the
5
values by using the hidden button to select the
value to change, then use the other two buttons
to Decrease (LEFT) or Increase (RIGHT)the value.
Hidden
7
Now Use the
Right Button
Trigger points and
Alarm
Left (hidden Button) selects Option
Right two buttons change value as in 5
above.
Options
to Scroll across to
setting
SG