OfiTE 170-182-1 Instruction Manual

Hthp filter press for cement testing with 175-ml, double-capped test cell and n2 pressuring manifold, 115 v/230 v

Advertisement

Quick Links

HTHP Filter Press for Cement Testing
with 175-mL, Double-Capped Test Cell and N
11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com
#170-182: (115 V)
#170-182-1: (230 V)
Instruction Manual
Updated 6/20/2017
OFI Testing Equipment, Inc.
Copyright OFITE 2015
©
2
Ver. 3.0
Pressuring Manifold

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 170-182-1 and is the answer not in the manual?

Questions and answers

Summary of Contents for OfiTE 170-182-1

  • Page 1 175-mL, Double-Capped Test Cell and N Pressuring Manifold #170-182: (115 V) #170-182-1: (230 V) Instruction Manual Updated 6/20/2017 Ver. 3.0 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A. Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com Copyright OFITE 2015 ©...
  • Page 2: Table Of Contents

    Safety ....................8 Quick Start ..................12 Operation..................15 Data ....................20 Cell Cap Assembly ...............21 Maintenance ..................22 Diagrams ..................24 Test Cell ..................25 Warranty and Return Policy ............26 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 3: Intro

    175 mL units to a useful working temperature of 300°F (149°C). Anyone running tests above 350°F (177°C) must substitute a com- plete set of o-rings after each and every test. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 4: Specifications

    Test Cell Capacity: 175 mL Receiver Volume: 15 mL Heater: 400 Watt Power Requirement: 115 VAC, 5 Amps, 50/60 Hz 230 VAC, 3 Amps, 50/60 Hz OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 5: Components

    O-ring for Valve Stem, Viton 90D, Qty: 4 #171-190-058 O-ring for Rupture Disk, Viton 90D, Qty: 1 #171-190-060 O-ring for Cell, Viton 90D, Qty: 4 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 6: 170-20 Manifold Block

    Valve Stem O-ring #170-18 Cement Screen #171-190-057 O-ring for Valve Stem, Viton 90D #171-190-060 O-ring for Test Cell, Viton ® #171-23-1 Safety Pin with Lanyard OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 7 Interval Timer, 60 minute (#170-33) (#155-20) (Set Screw Cell Assemblies Only) HTHP Pressure Relief Tool Cell Carrying Tool (#170-40) (#170-91) (To release trapped pressure) Safety Shield (#171-06) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 8 Thermocouple Assembly Stand for HTHP Cell Assembly (#171-45-1) (#171-190-028) (Direct temperature measurement Of the fluid Inside the Cell) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 9: Safety

    HTHP Filter Press and especially when one is going to extreme temperatures and pressures. Above 350°F (176°C) All o-rings must be re- placed after each and every test. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 10 The temperature of the fluid inside the cell however will not be at the set tem- perature, so always allow one hour of heating time for the fluid, after the cell has been fully inserted into the heating jacket. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 11 HTHP filtration. Under high tempera- tures and pressures, Nitrous Oxide can detonate in the presence of grease, oil or carbonaceous material. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 12 1" (2.5 cm) 115 mL Not Recommended- Oil-based drilling fluid. > 350°F (176°C) Use 500 mL Cell Void Space in a 1,500 PSI rated cell body OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 13: Quick Start

    21. Connect the back pressure receiver to the bottom valve stem and secure it with the safety pin. 22. Connect a Nitrogen cylinder to the dual Nitrogen manifold. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 14 36. Unscrew the outlet cell cap. 37. Wash the filter cake and record the thickness to the nearest 1⁄32". 38. Clean and dry all components. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 15 (#171-10) Nitrogen Manifold (#171-24) Regulators Nitrogen Connection Manifold Block (For Inlet Pressure) Male End of (#170-20) Hose (For Back Pressure (⅛") Nitrogen Manifold (#171-24) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 16: Operation

    7. Make sure all threads are clean and free of debris. 8. Apply anti seize compound (#165-44-2) to the threads of the cell locking rings and valve stems. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 17 19. Move the thermometer from the heating jacket to the hole in the test cell. Allow the cell to heat for one hour. Start a 60 minute timer now. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 18 Recommended Minimum Back Pressure Test Temperature Vapor Pressure Minimum Back Pressure °F °C 14.7 1,104 1,704 1,898 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 19 The heating jacket will still be very hot. Be careful not to touch it while removing the back pressure receiver. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 20: Cement Screen

    41. Clean and dry the apparatus thoroughly after each use. Inspect all of the o-rings and replace any that show signs of wear or damage. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 21: Data

    “zero mark” at the very edge of the ruler is useful here. Cake descrip- tions may be subjective and such notations such as hard, soft, rubbery, and fine, etc. convey adequate information on cake quality. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 22: Cell Cap Assembly

    (#171-190-040) and screw it into the port. Locking Ring Retaining Ring Rupture Disk (#171-190-023) (#130-81-040) (#171-190-027) Rupture Disk Retaining Ring Locking Ring O-ring OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 23: Maintenance

    Replace the thrust plate with the curved edges downward. Place a new rubber diaphragm inside the spring case. m. Replace the spring case and hand tighten. #143-07 with #143-00-1 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 24 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 25: Diagrams

    Test Cell O-ring Cell Cap, Outlet (#171-190-060) (#171-190-032-S) Outlet Valve Stem (#170-16) Safety Pin (#171-23-1) With Lanyard Back Pressure Receiver (#170-06-1) Needle Valve (#170-32) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 26: Test Cell

    O-ring, #228 Viton, 90D (#171-190-060) Detachable Screen Mesh, (#170-18) O-ring, #228 Viton, 90D (#171-190-060) Outlet End Cap (#171-190-032-S) Retaining Ring (#130-81-040) Cap Screw (#171-190-023) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 27: Warranty And Return Policy

    In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided. OFITE makes no other warranties or guarantees to the buyer, either express or implied, and the warranties...

This manual is also suitable for:

170-182