Preventive Maintenance - EDI VERIS EMS IV Installation & Service Manual

Stimulators
Table of Contents

Advertisement

VERIS
Service Manual
11.

Preventive Maintenance

The only areas accessible for field service are external. Internal adjustments of the optics require
special tools and must be performed by qualified engineers at the factory.
Do NOT open the stimulator housing for any reason. There are no user-serviceable
parts inside.
User Maintenance
1. Painted external surfaces of the stimulators may be cleaned with a soft cotton cloth or
standard alcohol pads.
2. Avoid touching the large eye lens. Apply lens cover when the instrument is not in use.
If necessary, clean lens with eyeglass cleaning materials and soft cotton or micropore
cloth.
3. Calibration
Perform Grid Calibration and Auto-Calibration at periodic intervals of less than three
months and after updates to the VERIS software. Instructions are found below and in
the VERIS User Manuals.
4. VERIS
application software updates are provided from the EDI website free of charge
within the first year. Instructions for updating the software are provided in the "Release
Notes" file included with the updated software. After one year, an annual service fee is
charged. A surcharge may be added and a new key file may be required for major
updates. Contact EDI for details.
VG-119-D / 21-Jan-2017
EMS & FMS Stimulators
Page 12 of 16