WARNING: Do not connect additional multiple outlet devices or extension cords to the system. WARNING: Do not connect any device(s) not approved by EDI to the multiple socket outlets of the isolation power supply. WARNING: Unauthorized modifications to the VERIS hardware and / or software void the warranty and may increase the risk of injury to patient and/or operator.
Page 4
™ VERIS Service Manual EMS & FMS Stimulators CAUTION: Carefully adhere to best medical practices whenever administering medication for mydriasis and/or corneal anesthesia. These include proper instructions to the patient, administration of eye medication, additional care in the event of adverse reactions, and provision of and/or use of dark glasses after testing.
™ VERIS Service Manual EMS & FMS Stimulators Labels Component Labels 2.1. Electro-Diagnostic Imaging, Inc. Electro-Diagnostic Imaging, Inc. Electro-Diagnostic Imaging, Inc. 1551 McCarthy Blvd., 1551 McCarthy Blvd., 1551 McCarthy Blvd., Milpitas, CA 95035 Milpitas, CA 95035 Milpitas, CA 95035 Made in USA Made in USA Made in USA Electro-Diagnostic Imaging, Inc.
™ VERIS Service Manual EMS & FMS Stimulators User Profile Users of the system include physicians, scientists, and other medical or scientific personnel trained or otherwise experienced in visual electrophysiological testing. Product Lifetime: The product lifetime is 5 years Components of the FMS IV / EMS IV Stimulators ...
Page 7
™ VERIS Service Manual EMS & FMS Stimulators Power Consumption (at max luminance) +12 VDC 1.5 A +24 VDC 0.5A Performance and Operating Specifications Internal high-resolution, color, LCOS stimulus display 1280 x 1024 pixels at 60 Hz, 75 Hz, or 85 Hz ...
Carefully unpack the FMS IV/EMS shipping carton and the carton containing the light articulated arm. If either box is damaged, notify the carrier and EDI immediately. Check the content of each box against its packing slip when it is opened to account for all items.
Page 9
FMS IV. The other end is connected to the VGA connector of the switchbox. Older switchboxes that lack a VGA port require a Cable 76 having a DIN connector and RCA plug in place of one VGA connector. Call EDI for assistance in this situation.
™ VERIS Service Manual EMS & FMS Stimulators Stimulus Luminance Calibration (FMS IV and EMS IV Stimulators) Luminance calibration of the EMS IV/ FMS IV microdisplay requires a calibrated Spot Calibrator with Eye Camera Calibrator Adapter (ECCA). These components are provided with every Veris™ platform.
IR fundus illumination back on. This calibration will be correct for any multifocal stimulus array and for any refractive correction. EDI recommends repeating this procedure every 3-4 months or whenever a new version of the software has been installed. VG-119-D / 21-Jan-2017...
™ 4. VERIS application software updates are provided from the EDI website free of charge within the first year. Instructions for updating the software are provided in the "Release Notes" file included with the updated software. After one year, an annual service fee is charged.
There are no field serviceable parts within any of the stimulators. See "Troubleshooting" below in the event of a malfunction. If problem persists, contact EDI. Do NOT open the stimulator housing for any reason. There are no user-serviceable parts inside.
1 Keyspan USB adapter Keyspan Model # USA-28XG Any Vendor 1 EDI Serial Cable Cable-34-01 * Part pricing varies. Contact EDI for current pricing information. FMS IV on Arm EMS IV on Arm EMS/FMS Controller Keyspan USB Adapter Grid Calibrator...
™ VERIS Service Manual EMS & FMS Stimulators 15. Software ™ All stimulator functions are controlled by the VERIS application software. VG-119-D / 21-Jan-2017 Page 16 of 16...