TEC:ENABle:COND?
Parameters
Notes
Examples
46
LDC-3916558
C O M M A N D R E F E R E N C E
LDC-3916558 Device-Dependent Commands
The TEC:ENABle:COND? query returns the value of the selected module's TEC
Condition Status Enable register.
:
None. The response is the sum of the following:
1
Current Limit
2
Voltage Limit
4
N/A
8
Temperature Limit
16
N/A
32
N/A
64
Sensor Open
128
Open Circuit
The selected module's enabled TEC conditions can be set by using the
TEC:ENABle:COND command. The TEC condition status can be monitored by the
TEC:COND? query.
"TEC:ENAB:COND?" - response: 640, means that only the following two conditions will
be reported (in summarized form) to the 3916's Status Byte: In Tolerance (512) and Open
Circuit (128).
:
TEC
ENAB
F
P
RONT
ANEL
R
EMOTE
:
COND
256
N/A
512
In Tolerance
1024
Output On
2048
N/A
4096
N/A
8192
N/A
16384
N/A
32768
N/A
?