General
Size:
Weight:
Enclosure Material:
Mounting:
Operating Limits:
Clock Accuracy:
Battery:
Battery Life:
Memory Size:
Memory Usage:
Sample Intervals:
Sampling Mode:
Resolution:
Internal Temperature Sensor
Type:
Range:
Accuracy:
Calibration:
Equation:
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
25.4 mm x 58.4 mm x 81.3 mm (1" x 2.3" x 3.2")
74 g (2.6 oz)
Cycoloy® (GE Plastics) C120 ABS + Polycarbonate
Magnetic backing and security loop
-40 to 70°C (-40 to 158°F) and 0 to 100% Relative Humidity
(non-condensing)
±2 seconds/day
3.6 Volt Lithium, 1 Amp-hour
5 years under normal use (factory replaceable)
64KB (up to 32,640 readings per channel)
1. Continuous (First-in, First-out)
2. Stop when full (Fill-then-stop)
User-selectable rates from every 8 seconds to 34 minutes.
1. Average (over sample interval except for 8 second
interval)
2. Spot
8-bit (1 part in 256)
CMOSens® (by Sensiron)
-40 to 70°C (-40 to 158°F)
± 0.6°C @ 25°C (± 1.0°F @ 77°F)
± 2.0°C from -40 to 70°C (± 3.6°F from -40 to 158°F)
Factory calibration verification and NIST certificates are
available upon special request.
Use Equation [91] TRH 1000 Temperature in TrendReader
software. Other equations for this thermistor type may be
selected but will give incorrect results. You may also create
your own equations, please refer to the Equations section or
contact
Customer Service
TRH Data Loggers
at
ACR
Systems.