The applicable electrical safety regulations must be closely followed in all installation work! Before any welding works in the vicinity of the devices, make sure that operating voltage is not connected! Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
A materials list is delivered with the device. Upon request, the manufac- turer will provide more detailed instructions for recycling and disposal. Used devices may also be returned to the manufacturer for recycling and disposal against a separate fee. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
Fig. 2. Valmet MCA-F. pulp type-independence, insensitivity to flow speed, and single-point calibration. There are two different models of Valmet MCA sensors: - Valmet MCA-F Fork sensor (figure 2). - Valmet MCA-FT Flow-through sensors (figure 3). Valmet MCA also includes a user interface known as Transmitter Central Unit (TCU) (figure 1).
2. Structure Valmet MCA comprises the sensor unit and a user in- 2.2. MCA-FT terface called Transmitter Central Unit. There are two The body of the Flow-Through sensor is a pipe, which, sensor models: MCA-F Fork Sensor and MCA-FT Flow-...
2.3. Sensor Electronics 2.4. Transmitter Central Unit (TCU) Sensor electronics is the same for both Fork and TCU is Valmet MCA's user interface and calculator. Flowthrough sensors. Refer to figure 6. The sensor Perform operations using the number buttons and other electronics card is installed between the round cover buttons and the four-line display.
– After the sensor there should be a straight section of pipe, length at least 2 times the pipe diameter, before a change in pipe profile. – Adjacent to Valmet MCA-F sensor or one meter be- fore it there must not be objects inserted into the pipe.
Page 14
Fig. 1. Valmet MCA-F Installation on a vertical pipe: A. after a pump, B. after a horizontal pipe. Fig. 2. Valmet MCA-F Installation on a horizontal pipe: A. after a vertical pipe, B. after a bend in the pipe. Valmet Mircowave Consistency Transmitter...
Page 15
Fig. 3. Valmet MCA-FT Installation on a vertical pipe: A. after a pump, B. after a horizontal pipe. Fig. 4. Valmet MCA-FT Installation on a horizontal pipe: A. after a bend in the pipe, B. after a vertical pipe. Valmet Mircowave Consistency Transmitter...
Page 16
Fig. 6. Valmet MCA-FT installation directly after a pump. Fig. 5. Valmet MCA-FT installation directly after a pump. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
Valmet MCA-F Installing the process coupling: – Make a 71 mm hole in the flow pipe in line with the The fork-type Valmet MCA-F sensor is installed in a sensor. process coupling welded onto the side of the process – If necessary, shape the process coupling (curvature...
NOTE: Do not place the sensor cable on cable shelves that contain cables of motors, pumps or other electrical cables. 232 (9.13") 133 (5.24") 31 (1.22") 70 (2.76") Fig. 15. TCU shield mounting dimensions. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
Page 21
1000 or CS2+. – Connect the current output line to connector CSIN or CSIN2. NOTE: Valmet MCA's current supply is passive and needs an external source of current. The minimum supply voltage is dependent on resistance. Operating region Fig. 16. TCU's Current output load capacity.
Scale the current output according to Chapter 6.2. Calibrate the consistency according to Chapters 7.1 and 7.2. After you have completed these procedures, Valmet MCA is ready to measure consistency. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
5. User Interface and Operation 5.1. Transmitter Central Unit Forward arrow: Use the forward arrow in Edit mode The operation unit of Valmet MCA is the Transmitter to move to the next character. Central Unit (TCU). Down arrow: Use the down arrow to scroll down if Number buttons: Use number keys to enter numerical there are more than four lines.
Sample history Output signal 1 Output signal 2 User settings Device info Set clock Address Error log Diag.values Diag.alarms Chemical compensation Temperature compensation Sensitivity correction Recipes Fig. 2. Operations Menu Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
Use the up arrow and down arrow buttons to make your selection. If you Select ON, program make measure- ment signal level adjustment automatically when it goes too low. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
H W V e r : 0 0 0 1 S W V e r : 0 0 0 1 6.6. Address Valmet MCA's address on HART bus. S/N: Sensor electronics serial number. HW Ver: Sensor electronics version number. SW Ver: Sensor electronics software version number.
7. Calibration The Valmet MCA is delivered with factory calibration, The sample-taking display is as follows: which means that it will measure consistency as soon as you switch the power on. Factory calibration is per- P r e s s s a m p l e - k e y formed on the device in connection with final testing s t a r t s a m p l i n g .
Offset field. Enter the desired offset correction laboratory consistency. If the pulp contains fillers, enter value. If, for example, Valmet MCA shows a value that in the fourth row in the Filler field the filler amount, according to laboratory monitoring is 0.1% too high, which shows the proportional amount of filler as a per- enter an offset correction value of -0.1%.
L a b = 3 . 2 0 C s = 3 . 0 0 read Valmet MCA values for a certain sample. To ac- C u m . o f f s e t = - 0 0 . 0 0 % cess the sample history table, select Calibration / T = 4 7 .
Even if the effect of the limit on the current signal is blocked by an OFF setting, an error message still ap- pears on the main display status line and error data is entered into the error table. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
Using Chemical Compensation Large fluctuations in chemical content may cause errors After you enter the laboratory values, go to Special in Valmet MCA consistency measurements. These er- Functions menu and select 1 (Chemical Compensa- rors can be compensated for with chemical compensa- tion).
Page 36
Press the Edit button. Then enter the correction temperature area must be sufficiently broad. curve as pairs of points. Valmet MCA then performs The results of the laboratory monitoring are shown the correction based on lines drawn through these on an MCA Laboratory vs.
Page 37
The effect of a new consistency calibration Changing the correction If a new consistency calibration is performed on Valmet If after the correction there is still some amount of MCA by taking a new sample after temperature consist- temperature dependency evident, you can most easily ency correction is determined, the measurement pro- perform the correction by editing the existing points.
9.2. Sensitivity Correction The sensitivity of Valmet MCA consistency measure- To go to the Sensitivity Correction menu, select ment is set according to the wood fiber. If the device Special Functions/Sensitivity Corr. The following display measures something else, the sensitivity may have to appears: be changed.
If necessary, scale the current output according to the instructions in section 6.2 and configure the special functions according to the instructions in chapter 9. Configure recipes 3 and 4 in the same way. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
Page 40
The text ‘ON’ next to the recipe number indicates which recipe is selected. Press Enter to clear the recipe. Note: You cannot empty the recipe that is currently in BIN0 BIN1 Recipe use. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
Valmet MCA does not require scheduled maintenance. is not able to escape properly from the pulp before The instructions in this chapter refer to fault conditions.
Page 42
Checking for antenna contamination requires a formed by Valmet maintenance personnel. break in the process and removal of the device. • Reference Channel Failure Refer to Sensor Electronics Failure.
Rlev: Shows the device's internal reference channel's signal level in millivolts. A decrease in the signal level to below alarm limits is always caused by a sensor electronics failure. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
- TCU: if you did not replace the TCU. - Default calibration: if you have replaced the TCU as well. Fig. 1. Removal diagram for sensor electronics. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
100 cable to the side so that the connector is able to be removed. 14. Pull the microwave cable carefully out through the feed-through hole on the shaft toward the flush- mounted antenna. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
Page 47
14. Install the sensor electronics as described in Chapter 11.1. Fig. 3. Removal diagram for antenna cables of other FT Sensor models. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
NOTE: Lift the cable by its casing tube, not by the cable itself. NOTE: Be careful not to dent the antenna cable connector on the base plate. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
Screw the probe antenna into the sensor body using socket key OUL00325 to 10 Nm torque. Install the antenna cable(s) as described in Chapter 11.2. Attach the sensor electronics as described in Chapter 11.1. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
NOTE: The outlet pipe releases hot air. Check that there are no cables that touch the outlet pipe or in its near vi- cinity. Fig. 4. Installing Vortex Cooler. Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
MCA, the following menu appears. program version C or higher and a HART Communic- ator that contains the Valmet MCA command base. HART Communicator is connected in parallel to the M C A : M C A - 1 2 3 4 current output, e.g.
12.2. Measurement 12.3. Configure Measurement menu shows the Valmet MCA measure- In Configure menu you can configure Valmet MCA's ment results. current output and certain other parameters. Configure menu also shows device information. MCA: MCA-1234 Measurement M C A : M C A - 1 2 3 4...
Press OK to start sample taking Sensor SW rev: Sensor electronics software revision. ABORT OK Press F4 (OK). Valmet MCA begins to average the S/N A: First part of TCU serial number. measurement result for the sample and the following display appears: S/N B: End part of TCU serial number.
Page 54
HART Communicator sends the value to Valmet MCA. Lab Cond Filler Enter the necessary laboratory information: Valmet MCA measures filler, but at a slightly different Lab Cs: Laboratory consistency. sensitivity than fibers. The Filler function allows you to inform the device about filler changes if the relative...
Page 55
Large fluctuations in chemical content may cause errors M C A : M C A - 1 2 3 4 in Valmet MCA consistency measurements. These er- 2 3 - 0 4 - 0 5 1 0 : 3 2 : 0 9 L a b = 3 .
Page 56
Recipes You can define temperature compensation correction The Recipe function may be necessary in special curve for Valmet MCA if you notice temperature depend- functions in which measurement conditions change so ency in the measurement results. much that one calibration cannot cover it.
Diag History: The 10 most recent self-diagnostics alarms with beginning and ending times. Status, Status nxt: State of diagnostics variables (alarm on/off). Diag values: Measured values of diagnostics variables (see Chapter 8.2). Valmet Mircowave Consistency Transmitter Installation & Owner’s manual OUL00298 V2.3 EN...
1 (1) Valmet MCA Technical specification Valmet MCA sensor Operating terminal TCU Measurement Connections Cable to sensor .... length 10 m (33 ft), Measuring range ... 0–16 % Cs; if greater than extension cable 10 m (33 ft) 16 %Cs or if used in an available, max.
Page 61
1 (4) Valmet MCA Service kits & spare parts Valmet MCA Service Kit OUL00335 Order Code Description OUL00347 Accessory Case OUL00298 Manual OUL00051 Sensor Electronics OUL00241 Operating Unit TCU OUL00334 Simulation Cable OUL00325 Tool for Sensor Head OUL00323 Tool for Fork Pt-100 Cap...
Page 65
Dispositivo Valmet Microwave Consistency Laite Dispositivo Enheten Dispositivo Transmitter (Valmet MCA) Gerät Urządzenie Artifice This product is in conformity with the requirements of the following Este producto es conforme a los requisitos de las siguientes directives, and with standards and national legislation directivas y normas y legislación nacional de aplicación de estas...
Page 66
This document is the exclusive intellectual property of Valmet Corporation and/or its subsidiaries (collectively “Valmet”) and is furnished solely for operating and maintaining the supplied equipment and/or software. Use of the document for any other project or purpose is prohibited. All copyrights to the document are reserved by Valmet.
Need help?
Do you have a question about the Valmet MCA and is the answer not in the manual?
Questions and answers