Crowcon C-Test User Manual

Gas (bump) test and calibration station for use with the crowcon clip and crowcon clip+

Advertisement

Quick Links

Crowcon C-Test User Manual
Crowcon C-Test
Gas (Bump) Test and Calibration Station
For use with the Crowcon Clip and Clip+
Installation and Operation Manual
Crowcon contacts:
UK:
2 Blacklands Way, Abingdon Business Park, Abingdon, OXON, OX14 1DY
+44 (0) 1235 557700 sales@crowcon.com
US:
21 Kenton Lands Road, Erlanger, Kentucky 41018-1845
+1 859 957 1039 salesusa@crowcon.com
NL:
Vlambloem 129, 3068JG, Rotterdam
+31 10 421 1232 eu@crowcon.com
SG:
Block 194 Pandan Loop, #06-20 Pantech Industrial Complex, Singapore, 128383 +65 6745
2936 sales@crowcon.com.sg
CN:
Unit 316, Area 1, Tower B, Chuangxin Building, 12 Hongda North Road, Beijing
Economic & Technological Development Area, Beijing, PRC 100176
saleschina@crowcon.com
+86 10 6787 0335
www.crowcon.com
/
Issue 1 - June 2012
www.crowcon.us
Page 1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the C-Test and is the answer not in the manual?

Questions and answers

Summary of Contents for Crowcon C-Test

  • Page 1 Crowcon C-Test User Manual Crowcon C-Test Gas (Bump) Test and Calibration Station For use with the Crowcon Clip and Clip+ Installation and Operation Manual Crowcon contacts: 2 Blacklands Way, Abingdon Business Park, Abingdon, OXON, OX14 1DY +44 (0) 1235 557700 sales@crowcon.com 21 Kenton Lands Road, Erlanger, Kentucky 41018-1845 +1 859 957 1039 salesusa@crowcon.com...
  • Page 2: Table Of Contents

    Crowcon C-Test User Manual READ FIRST BEFORE OPERATION Do not use the C-Test if it appears to be damaged. Inspect prior to each use. Ensure the C-Test is used with certified calibration gas matching the configured concentrations. DO NOT use expired calibration gas. Please check the expiration date located on the bottle.
  • Page 3: Getting Started

    1. Getting Started 1.1. Set Up / Installation instructions C-Test is designed to be used on a flat surface and can be mounted on either a desk or on the wall. Each C-Test should come with the following: C-Test station...
  • Page 4: Indications

    2.3. Hibernation C-Test has the ability to hibernate the Crowcon Clip+ monitor. To use this feature, power on the C-Test, place the monitor in the bay, press and hold down the bump test button and calibration button at the same time for 2 seconds. All lights will turn orange and the power light will blink. Once complete, the C- Test lights will turn green for a successful hibernation and red for a failure.
  • Page 5: Orientation

    Power and Test LED’s Monitor Bay 5. Maintenance The C-Test Station should be examined before and after use to ensure that all components are in operating order and undamaged.  Ensure that the gas cylinder has adequate gas for testing purposes.
  • Page 6: Advanced

    C-Test Manager Software. Caution: Do not delete any files from the USB prior to saving them; the C-Test does not store this information internally outside of the USB.
  • Page 7  User ID: Unique programmable 6 digit number  Self Test Interval: Default is 20 hours; this option can be changed via IR Link or via C-Test Manager The Third line(s) are for each sensor on the monitor. ...
  • Page 8: C-Test Manager Software User Guide

    7.5. Getting started C-Test Manager interacts with the C-Test either via IR Link or by USB stick. The software allows the retrieval of test logs, Crowcon Clip & Crowcon Clip+ event logs, program configurations, and managers to perform firmware upgrades. Below are detailed instructions for how to use the software: 7.6.
  • Page 9: Changing C-Test Settings

    Auto cal after Bump fail – Allows a calibration to run should a monitor fail and bump test Firmware upgrade: The C-Test allows for firmware upgrades to occur during the lifetime of the unit. These may be product improvements to add additional features or to enhance monitor performance.
  • Page 10: Monitor Settings

    C-Test. Self Test Interval: By default all Crowcon Clip & Clip+ instruments have a default 20 hour self test. Frequency can be set anywhere from 8 to 20 hours.
  • Page 11: Log Settings

    Right click next to the “Files” tab, and then select Set Inventory Filename. A screen will prompt you to choose the location where you want to save the file and file name. Return back to the C-Test Manager and click on the box next to TS-CC _[Serial Number], then click “Read” again. Then go to the specified location to find the inventory file.
  • Page 12: Warranty

    Crowcon warrants this product to be free from defects in material and workmanship under normal use and service for a period of one year beginning upon the date of delivery for all Crowcon C-Test monitors. This warranty extends only to the sale of new and unused products to the original buyer.