CME T34L Operator's Manual

CME T34L Operator's Manual

Syringe pump

Advertisement

0344
Caesarea Medical Electronics Ltd.
16 Shacham street, North Industrial
Park Caesarea
P.O.Box 4294, Caesarea 38900, Israel
Tel: +972-46271737
www.cme-infusion.com
Software Version LCAT / LCATBOL
T34L
Syringe Pump
Operator Manual
MedNet GmbH.
Borkstrasse 10
48163 Muenster
Germany
T: +49 251 32266 0
100-090SL Rev. 03

Advertisement

Table of Contents
loading

Summary of Contents for CME T34L

  • Page 1 T34L ™ Syringe Pump Operator Manual 0344 Caesarea Medical Electronics Ltd. MedNet GmbH. 16 Shacham street, North Industrial Borkstrasse 10 Park Caesarea 48163 Muenster P.O.Box 4294, Caesarea 38900, Israel Germany Tel: +972-46271737 T: +49 251 32266 0 www.cme-infusion.com Software Version LCAT / LCATBOL...
  • Page 2 It is also recommended that before being allowed to use the device all personnel should receive training from either Caesarea Medical Electronics Ltd. (CME) or those trained by Caesarea Medical Electronics Ltd. (CME) to deliver training on this device. Competency of all attendees should be assessed with both...
  • Page 3: Table Of Contents

    12 – Alarm Conditions and Troubleshooting Alarm Conditions ..................... 22 Troubleshooting ....................... 23 13 - Change Set Up 14 – Specifications 15 – Service and Maintenance Cleaning ........................26 Storage ........................26 Service ........................26 16 – Default Syringes LIMITED WARRANTY T34L™ Operator Manual...
  • Page 4: Introduction

    Warnings, Cautions & Notes as described below. This manual will ensure the user will: 1. Be able to identify the key features of the T34L and understand their function 2. Be able to correctly load a syringe into the T34LTM 3.
  • Page 5: Modes Of Operation

    NOTE: your T34L can be configured to one of the above modes (as appropriate to your clinic/setting) during installation by either CME or your service centre. Take time to consider which mode is best for your practice. T34L™ Operator Manual...
  • Page 6: Keypad Description And Functions

    Forward key YES/START key Confirms choices Minus key scrolls between options Backward key a. shows technical data and Battery status Info key b. Long press will Lock/Unlock Keypad Pump display Backlight can be enabled when any key pressed T34L™ Operator Manual...
  • Page 7: Symbols, Warnings And Cautions

    4 – Symbols, Warnings and Cautions System Symbols The following symbols are used on the T34L Syringe Pump and components. Labels on the pump or statements in this manual preceded by any of the following words and/or symbols are of special significance and/or are intended to help you to operate the pump in a safe and successful manner.
  • Page 8: Terms Used In Manual

    Use in ambulatory regimens by patients who do not possess the mental, physical or emotional capability to self-administer their therapy; or who are not under the care of a responsible individual Warnings To avoid the possibility of injury or death, please note the following advice before using the T34L Syringe Pump: Warning: CME syringe pumps should not be used for infusion of medications whose stoppage or interruption could cause serious injury or death.
  • Page 9: Cautions

    The equipment is not suitable for use in the presence of flammable mixtures. Warning: syringe pump must always be used in a CME Ltd. pouch or similar T34L receptacle if used in direct sunlight. If the pump is exposed to direct sunlight, it may affect functionality.
  • Page 10: Installation And Set Up

    Remove the T34L Syringe Pump & accessories from the packaging and inspect for damage during shipment or storage. Warning: Do not use T34L if there are any obvious signs of damage. Return for inspection by authorized service personnel. Make sure you have the following items: ...
  • Page 11 Battery Level, in all cases verify sufficient battery charge is available to complete the current program. If not, charge the battery. Empty Full  Sufficient battery charge may vary depending on your setting, infusion rate, number of times the information options/screen backlight are used. T34L™ Operator Manual...
  • Page 12: Administration Sets

    6 - Administration Sets The T34L Syringe Pump System can be operated with any extension set with a luer connection however, for system optimization, accuracy & performance, CME recommends the use of CME proprietary sets. Proprietary sets available from CME include:...
  • Page 13: Access Codes & Keypad Lock

    In normal use the nurse will not see these fields or be prompted for the technician codes. Keypad Lock As well as the ability to lock or limit certain infusion parameters or pump settings, T34L allows users to lock the operation of the keypad if concerned about patients, relatives or un-trained personnel tampering with the pump.
  • Page 14: Infusion Set Up & Programming

    During Pre-Loading the actuator returns to the start position of the last infusion programmed. If the user regularly uses the same syringe brand, size and fills to the same volume the actuator will return to the correct position each time & not require adjustment. T34L™ Operator Manual...
  • Page 15: Syringe Detection & Confirmation

    Ensure the patient is NOT connected to the set & Press Yes to confirm. c. Press and Hold the FF key until the slack is removed and/or purge is completed d. Press STOP to return to programming screens when purge is completed. T34L™ Operator Manual...
  • Page 16: Setting Infusion Parameters

    10ml BD Plastipak 17. The Pump calculates and displays the rate (in milliliters per hour) required to deliver the VTBI over the infusion duration confirmed. Rate 0.5 ml/h Confirm, Press YES 18. Press YES to confirm the calculated rate. T34L™ Operator Manual...
  • Page 17 NOTE: In VOLUME OVER TIME Senior clinical personnel or technicians may have set a MAXIMUM RATE LIMIT to prevent T34L being programmed at rates above a fixed limit in line with the application the pump is regularly used for. Pump will not allow the user to scroll to a duration that requires a rate above this limit to deliver the contents of the syringe over that time.
  • Page 18: Accessories

    The pouch will also protect the pump from damage or syringe displacement. When using the CME Ltd re-useable (washable) pouch, it is possible to access the screen and keypad of the device during infusion by lifting the Velcro flap of the pouch whilst the pump remains in the pouch.
  • Page 19: Pole Mounting Unit

    Pole Mounting Unit The T34L is pole mountable, vertically of horizontally. Connect the pole mounting unit to the pole using the adjusting screw, and just click in the T34L back hooks onto the pole mounting unit. Mounting device for two syringe pumps is also available.
  • Page 20: Recommended Checks During Infusion

    Travel 3.61 mm/ml event.  NOTE: Each event is assigned a number and the pump stores the last 512 in memory. The event log “loops” adding the newest and deleting the oldest event. T34L™ Operator Manual...
  • Page 21: Advanced Functions

    Maximum Rate limit from the Technician menu. For example senior clinicians could decide to limit the maximum possible rate programmable on the pump to 3ml/hr for both rate and titration limit to prevent accidental over infusions. T34L™ Operator Manual...
  • Page 22: Alarm Conditions And Troubleshooting

    Occlusion alarm will be operated after 5 hours on minimum rate of 0.05 ml/hr, after one minute on rate of 10 ml/hr and after less than 5 seconds on maximum rate of 1500 ml/hr (using one meter tubing inner diameter 0.5 mm). T34L™ Operator Manual...
  • Page 23: Troubleshooting

    Pump faulty or incorrectly Service/calibration required calibrated. The Pump has stopped Exhausted Charge battery, turn pump before empting the syringe on, confirm syringe size and brand, select resume to continue infusion. Blocked/trapped infusion set Clear the occlusion T34L™ Operator Manual...
  • Page 24: Change Set Up

    Activates Keep Vein Open infusion at End KVO Operation Rate Program 2.0ml/h Prevents alteration of default duration and infusion Program Lock rate during set up (with Program Lock ON Default ON/OFF Duration cannot be set to zero) T34L™ Operator Manual...
  • Page 25: Specifications

    Indicators: Graphic Display 128 x 64 pixels with back light (8 keys) & dual color operation LED Alarms: When a problem is detected, the T34L displays the following alarm messages, sounds an audible alarm and the red LED lights: ...
  • Page 26: Service And Maintenance

    These chemicals can cause damage to components and labels. Use a lint-free cloth dampened with warm water and mild detergent or 10% bleach solution. Caution: Do not soak or immerse any part of the T34L in water or any other solution. Storage If the pump is to be stored for an extended period it should be cleaned.
  • Page 27: Default Syringes

    Time to Alarm from Occlusion The Time to Alarm from Occlusion for the T34L Syringe Pump are in accordance to the table below. All the tests were performed using a CME extension set, 100-172S and a BD Plastipak syringes filled with water.
  • Page 28: Limited Warranty

    This Limited Warranty is the sole and entire warranty pertaining to CME’s products and is in lieu of and excludes all other warranties of any nature whatsoever, whether stated, or implied or arising by operation of law, trade, usage or course of dealing, including but not limited to, warranties of merchantability and warranties of fitness for a particular purpose.