Instructions for Operation and Care of Forma-Splint Thermal Bath
This Whitehall product has been carefully packaged at the factory to minimize the possibility of damage during shipping.
— Inspect the packaging for external signs of
— Inspect the contents for damage.
If there is visible damage to the instrument upon receipt, inform
the shipping company and Whitehall immediately.

SET-UP INSTRUCTIONS

WARNING
!
Do not attempt to operate this equipment if there is
evidence of shipping damage or you suspect the unit
is damaged. Damaged equipment may present
additional hazards to you. Contact Whitehall technical
support for advice before attempting to plug in and
operate damaged equipment.
1. Your new Splint Pan includes the following:
(1) Heating Appliance
(1) Power Cordset
(1) Warranty Registration Card
(1) Operation & Maintenance Manual
NOTE: If you are missing any of the above,
contact WHITEHALL MANUFACTURING
toll free telephone no. (800) 782-7706 or (626) 968-6681
2. Unpacking your Splint Pan
Remove all shipping & packaging material. Remove all
items from inside of tank such as Power Cordset, Filler
Hose, Warranty Registration Card, and Operation &
Maintenance Manual.
3. Location of your Splint Pan
Place the Splint Pan at a convenient height & location
such as on a table or scratch-resistant counter top.
IMPORTANT: Appliance must be placed on a level
surface! Splint Pan must be positioned so to allow ease of
disconnecting Power Cordset if necessary.
42
4
4. Preparing to Fill your Splint Pan
Make sure appliance is unplugged and all packaging has
been removed from reservoir. Ensure that the inside of
tank is thoroughly clean before filling with water.
DO NOT use any medium except water.
Check drain valve and make sure handle is in the "closed
position" by turning clockwise to make certain no water in
tank will drain out of appliance while being filled.
WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A
Phone (800) 782-7706 • (626) 968-6681 • Fax (626) 855-4862 • Web: www.whitehallmfg.com

INSPECTING THE SPLINT PAN

damage.

Inspection Checklist

Is the Splint Pan free of excessive wear?
Is a properly-grounded and voltage-matched
receptacle available?
Is the outlet equipped with a functioning GFI?
Does the drain valve open and close properly?
5. Filling your Splint Pan
SET-UP INSTRUCTIONS
(continued)
6. Preparing to turn "ON" your Splint Pan
Close lid.
Lid can pinch fingers.
Locate the Power Switch at front of the appliance and
make sure it is set to "OFF". Insert connector end of power
cordset into power inlet at back of appliance.
Connect plug end of power cordset into a G.F.C.I.
protected AC electrical receptacle.
7. Turning "ON" your Splint Pan
You are now ready to turn "ON" the appliance. Locate the
Power Switch at the front of the unit and press to "ON"
position.
8. Adjusting/Setting the Heat-Up Temperature
To change the set temperature on the model SP-1501,
drain the water and unplug the unit. Carefully remove the
four screws holding the nameplate and locate the two
thermostat knurled knobs. The top knob is the thermostat
for the high temperature limit setting and the bottom knob
is the thermostat for the low temperature limit setting.
Insert a small screwdriver into the slotted head of the
thermostat
knob
turning
clockwise
to
raise
the

Advertisement

Table of Contents
loading

Table of Contents