Summary of Contents for De Nora Capital Controls CHLORALERT T17CA4000 Series
Page 1
Instruction Manual CHLORALERT Series T17CA4000 Chlorine Gas Leak Detector CAPITAL CONTROLS ® - 1 - 325.6030.7...
Page 2
De Nora Water Technologies, Inc. cannot be responsible for the overall system design of which our equipment may be an integral part of or any unauthorized modifications to the equipment made by any party other that De Nora Water Technologies, Inc.
Page 3
Should any question arise which may not be answered specifically by these instructions, they should be directed to De Nora Water Technologies for further detailed information and technical assistance. DNWT takes all possible precautions in packing each equipment item to prevent damage during shipment.
INTRODUCTION General Description The De Nora Water Technologies chlorine gas leak detector, CHLORALERT™ T17CA4000, continually monitors air samples for the presence of chlorine. It is usually installed in chlorine usage and storage areas to protect personnel and equipment. A small internal air blower continually draws a sample of air into the Chloralert. The flowmeter and valve enable a fixed amount of air to be passed over the surface of the special electrolyte, where the chemical reaction takes place in presence of chlorine.
INSTALLATION Location The Chloralert™ should ideally be mounted outside the area to be monitored. However it can be mounted in the contaminated area if it is considered absolutely necessary. The location should be such that the ambient temperature never falls below -20 °C (-4°F) nor rises above 65 °C (149°F). It should never be mounted in direct sunlight and should be protected from the weather if mounted outside.
Electrical wiring Cable gland on the Chloralert™ will accept a cable of diameter between 5 to 10 mm. Refer to the following drawings to connect the power supply and the external signals to the Chloralert. Figure 4 - Electrical Wiring - 11 - 325.6030.7...
START-UP, OPERATION AND MAINTENANCE Electrolyte filling at start up Refer to figure 5. Switch off power supply or unplug instrument. Remove cover rising it. Disconnect cell cable from terminal block noting carefully position of each cable. Disconnect exhaust tubing from the cell. Unscrew the four retaining screws and drop bottom half of cell from the upper for filling.
Circuit performance check Simultaneously press and hold down both TEST and RESET switches. The alarm LED must change from the steady condition to flashing and the sampling fan stays activated. This checks that the electronic circuits are operating properly. If it is desired to check external circuits the RESET switch must be released: any remote device connected to the relay will be actuated.
MAINTENANCE PROCEDURES Note: disconnect power supply before removing the instrument cover. Measuring cell maintenance Refer to Section 5.1 and 5.2 of this manual. Flowmeter 6.2.1 For Model T17CA4100 Clean tube and float whenever dirt has accumulated causing malfunction. Using a screwdriver loosen and slide the top tube retainer ring towards the enter of the tube. Remove tube by pulling forward.
CE DECLARATION Chlorine Gas Leak Detector Model T17CA4000 is in conformity with the following standards: EN 56681-2 (Generic emission standard, industrial environment) • EN 56682-2 (Generic emission standard, industrial environment) • Following the provisions of directives: “Electromagnetic compatibility 89/336/CEE, modified by directive 92/31/CEE". - 15 - 325.6030.7...
Need help?
Do you have a question about the Capital Controls CHLORALERT T17CA4000 Series and is the answer not in the manual?
Questions and answers