4.2 Maintenance and Service Procedures ..........14 4.2.1 UPC5100/UPC5110 Panel/Chassis - Removal and Installation ......14 4.2.2 Nitrogen Cylinder Assembly (PN 59531) - Removal and Installation .
Page 3
Model Number System ......................36 Available Ranges, Conversions and Resolutions ..............37 Options - Replacement Kits ....................... 38 Specifications ..........................39 UPC5100/UPC5110 Warranty and Return Policy ................... 51 UPC5100/UPC5110 Return Material Authorization Form............... 52...
Introduction Utilizing microprocessor technology, the UPC5100 and rack-mounted UPC5110 instruments offer a combination of features, performance, versatility, and reliability not previously available in a single, self-contained pressure calibration instrument. Some of the features are listed below: •...
Page 5
The UPC5100 has an internal, rechargeable 12 volt lead acid battery, that provides a minimum of six hours of complete portability when fully charged. An ON/OFF and battery test switch is provided to conserve energy when the instrument is not in use and to provide the operator with battery voltage status during use.
Page 6
The following schematic provides an overview of the UPC5100/UPC5110’s function. Figure 1-1. UPC5100/UPC5110 Flow Diagram Introduction...
The following sections explain the various procedures for operating the UPC5100/UPC5110. Pressure Cylinder Filling Procedure To initially fill or refill the internal pressure cylinder (2015 PSI max) of the UPC5100/UPC5110, see Figure 2-1 and proceed by following these steps: NOTE: Check the Nitrogen Fill Port markings on Pressure Console, some units may be 2216 PSI max.
Figure 2-2. Initial Setup Procedure NOTE: UPC5100 shown, AC input (7) and fill port (4) are on back side of UPC5110 Rack Mountable Calibrator. 6. Optional - if the current (4.0000 to 20.000 mA) measurement features are used, connect the provided transducer test cable (PN 55092), to the transducer test jacks (14).
0-100.00 mV by 0.02 mV Table 2-1. Display Select Switch (16) NOTE: UPC5100/UPC5110 reads a 4-20 mA signal only, but will display as either 4-20 mA or 20-100 mV. The test cable connector wiring is as follows: CONNECTOR PIN DESIGNATION...
Figure 2-3. Pressure or Vacuum Measurement Sequence NOTE: UPC5100 shown, AC Input (7) and Fill Port (4) are on back side of UPC5110 Rack Mountable Calibrator. Vacuum Measurement Sequence See Figure 2-3 above and proceed as follows: 1. Turn the MODE SELECT valve (10) to the VACUUM position.
The UPC5100/UPC5110 can be operated and recharged by connecting to a standard AC outlet via the line cord (supplied). The battery re-charge cycle time is approximately 16 to 20 hours with the ON/OFF switch in the OFF position.
• The CONDEC Repair Lab is equipped to do calibrations on CONDEC calibrators and pressure standards. Calibrations include a certification and are traceable to N.I.S.T (see “UPC5100/UPC5110 Return Material Authorization Form” on page 52).
Zero/Span Calibration Selecting the position on the Condec Calibration Module (PN 60109) places the instrument into its ZERO/SPAN calibration mode. The display is shown in Figure 3-2. ZERO/SPAN Figure 3-2. Zero/Span calibration for Gage-Only Units The unit needs to be calibrated to only one positive and one bi-directional range. All bi-directional ranges are calibrated over 0 to +14 PSI and 0 to -14 PSI.
Install the Condec Calibration Module (PN 60109) and select the position of the rotary switch on the LYN/HYS module. This places the UPC5100/UPC5110 into its linearization/hysteresis calibration mode. The display is shown in Figure 3-3 below. NOTE: The zero/span calibration needs to be performed prior to linearity and hysteresis calibration.
If 100% (±0.05%) is not obtained, repeat the zero/span calibration sequence. Shunt Resistor Calibration To place the UPC5100/UPC5110 into shunt calibration mode, install the Condec Calibration Module (PN 60109) and select the position of the rotary switch. The display is shown in Figure 3-4.
NOTE: If the display reading is off, set the Current Generator to 0, and press the ENTER button on the Condec Calibration Module. Set the Current Generator for 20 mA output. The display will read 20.000. If the display reading is off, press the ENTER button on the module.
Maintenance and Service This section outlines the mechanical and basic electrical repair procedures for the UPC5100/UPC5110. Troubleshooting Use Table 4-1 below for information on troubleshooting the UPC5100/UPC5110. Symptom Problem Remedy No lit display Unit will not energize Check fuse, check power source, check...
Nitrogen Cylinder Assembly (PN 59531) - Removal and Installation NOTE: Condec strongly recommends that the internal nitrogen supply cylinder be pressure-tested and re-certified every five years from date cylinder was manufactured per U.S. DOT. 3AL Regulation, Title 49 CFR, parts 173 and 178.
5. To disassemble the isolation valves (inner valve), first remove the valve needle (18) by turning the gear (6) clockwise. 6. Loosen and remove the valve housings (19) using the isolation valve housing removal socket (68509) and socket wrench. UPC5100/UPC5110 Operation & Maintenance Manual...
7. Remove the valve stem seats (8) and valve needle seats (9) using the needle-nose pliers. 8. Remove the inner and outer O-rings (28, 27) and back-up rings (31, 30) from the valve stem seats and wash all parts in solvent (de-natured alcohol). 9.
14. Apply a small amount of Krytox grease to the threads of the valve needles (18) and install ISOLATION into the valve by turning counter-clockwise. Rotate the gear until the needle just stops at the seat. UPC5100/UPC5110 Operation & Maintenance Manual...
4.2.8 ORION-2D Manifold - Panel Installation Tools required: 7/16" open end wrench Phillips screwdriver hex wrench (.061") snoop, liquid leak gas detector (PN 64781) 11/32" open end wrench (thin) 1. If not already done, remove the panel knobs from the (pressure), , and valves...
Page 23
VENT knob insert (4). Do not use excessive torque when adjusting valve. The seat can be damaged. Caution 31. Tighten the set screws (25) with the .061" hex wrench. The valve is now adjusted. VENT UPC5100/UPC5110 Operation & Maintenance Manual...
MONITOR pull regulator knob outward and turn counter-clockwise. When closed, push knob inward. VENT valve should be left open. 2. Rotate SUPPLY SELECT to N position. 3. Turn customer supplied N source valve to the off position, vent pressure line, then install line to fill port of UPC5100/UPC5110. Maintenance and Service...
5. Fill the cylinder to approximately 1000 PSIG and check all fittings for leaks. If there are no leaks, fill nitrogen supply cylinder to maximum pressure. See Section 2.1 on page 4 for cylinder refilling procedure. 6. Install panel/chassis assembly in its enclosure as described in Section 4.2.1 on page 14. UPC5100/UPC5110 Operation & Maintenance Manual...
2. Remove front panel from its enclosure as described in Section 4.2.1 on page 14, and carefully place on a bench top. 3. Remove nylon tubing from the compressed air port (top/input) and the vacuum port (front/output) of the vacuum generator. 4. Using nutdriver remove nylon spacers, (PN 60472), and remove vacuum generator. UPC5100/UPC5110 Operation & Maintenance Manual...
Page 28
NOTE: Some vacuum generators are mounted with screws and nuts or adhesive. 5. Clean the vacuum generator by using compressed air. Do not use solvent for cleaning. Warning Removal and Cleaning - Aluminum Body (PN 54965) 1. Remove ORION-2D manifold from front panel as described in Section 4.2.3 on page 16, and carefully place on a bench top.
4. Turn the MODE SELECT knob on the front panel to VACUUM position. 5. Turn RANGE SELECT switch to +/- 28 in. Hg. 6. Energize UPC5100/UPC5110 and allow warmup time. 7. Push the ZERO button (less than 5 seconds), then close the VENT valve.
11. Install panel/chassis assembly in its enclosure as described in Section 4.2.1 on page 14 4.2.17 Inlet Check Valve, Nitrogen Fill Port (PN 60263) - Removal, Disassembly and Installation Remove the check valve if it does not hold the pressure of the N cylinder.
BATTERY TEST 3. Plug in the harness connector to its receptacle J7 on the CPU board and install cable clamp. 4. Install panel/chassis assembly in its enclosure as described in Section 4.2.1 on page 14. UPC5100/UPC5110 Operation & Maintenance Manual...
4.2.21 Power Switch (PN 58878) for Non-Battery Units - Removal and Installation Tools required: Phillips screwdriver 11/16" open end wrench A/R soldering iron A/R shrink sleeving (PN 64567) A/R heat gun Removal: 1. Disconnect the power cord from the power source and line filter. Remove front panel from its enclosure as described in Section 4.2.1 on page 14, and carefully set on a bench top.
3. Tighten the switch mounting nut and lock washer from the rear of panel. If wrench is used, do not over tighten, damage may occur to switch. Caution 4. Install panel/chassis assembly in its enclosure as described in Section 4.2.1 on page 14. UPC5100/UPC5110 Operation & Maintenance Manual...
4.2.24 Power Supply Assembly (Battery Units Only) - Removal and Installation 120 VAC input (PN 58727); 220 VAC input (PN 58733). Tools required: Phillips screwdriver Flat blade screwdriver (small) 11/32" open end wrench or nutdriver Removal: 1. Disconnect the power cord from the power source and line filter. Remove front panel from its enclosure, as described in Section 4.2.1 on page 14, and carefully set on a bench top.
Page 35
Adjustment - Battery Voltage Display Reading NOTE: The BATTERY TEST should only be performed with the UPC5100/UPC5110 operating at Zero pressure (VENT valve open) and at the conclusion of the test, the unit's ZERO button will have to be re-pushed to re-zero the instrument.
Page 37
O-RING, ITEM 28, .042 ID O-RING, ITEM 29, .799 ID O-RING, ITEM 32, .424 ID THESE COMPONENTS TO BE INSTALLED AT LOCATION THESE COMPONENTS TO BE INSTALLED “C” ONLY. AT LOCATION B Figure 4-1. ORION-2D, Exploded View UPC5100/UPC5110 Operation & Maintenance Manual...
C - AC Only (220 VAC)*** ≠ D - Battery Operation & 220 VAC ≠ +-------------------------------------------- --+ ≠ DISPLAY ≠ A – Light Emitting Diode (LED) B – liquid Crystal (LCD) *** UPC5100, Available as a Special UPC5100/UPC5110 Operation & Maintenance Manual...
Available Ranges, Conversions and Resolutions NOTE: Non-standard ranges are available by special order. Calibrator Conversion Factors: • kPa = PSI x 6.89476 • in Hg (0˚F) = PSI x 2.036 • Bar = PSI x 0.0689476 • Kg/cm = PSI x 0.070308 •...
Note: A small coating of Fluorinated Krytox grease, (PN 55593), should be applied to both sides of O-ring prior to installation. Pressure Trap (Data Sheet # 58621) - PN 58487 Note: For UPC5100/UPC5110 model’s prior to serial number A1300, approximate date prior to 7-9-1990 use Pressure Trap, PN 58492, Data Sheet # 58627. Battery Replacement Kit - PN 55354 Test Port (output) Attachment Swivel Fitting - PN 55291 Test Port (output) Quick-Disconnect Male Hose fitting - PN 60195...
Specifications Pressure Specifications: Fill Port: Pressure ranges: +- 0-28 in Hg Style: Quick-disconnect. +- 0-387.00 in H Pressure Rating: 3000 PSIG connected, 2000 PSIG disconnected + 0-2771.0 in H Material: 300 series stainless steel +- 0-96.00 kPa Test Port: + 0-100.00 PSI Available Pressure Pressure Rating: 2000 PSIG...
Page 43
Enamel paint, textured finish UPC5110 Case Dim’s: 19" wide x 8.1" deep x 10.5" high Color: Gray (Case dimensions excluding front handles) Control Panel: Material: Aluminum (5052-H32) Thickness: 0.125 in Finish: Gray enamel paint with black silkscreen nomenclature UPC5100/UPC5110 Operation & Maintenance Manual...
Page 44
R33 LOCATION. TWIST, DO NOT SOLDER, LEADS TOGETHER. LEAVE APPROXIMATELY 1/4" LEAD LENGTH ON EACH RESISTER. OBSERVE POLARITY OF ICs, CAPACITORS, DIODES, AND TRANSISTORS. NOTE TO CONDEC MANUFACTURING: ITEMS 11 & 31 REQUIRED FOR UPC5200 &UPC5210 UNITS. REMOVE 3- TEST PROCEDURE: (A)KUA7317 ITEM 11 AND SOLDER ITEM 31 ONLY PRIOR TO INSTALLING INTO UPC5000 OR UPC5010.
Page 45
Figure 8-2. CPU & POWER SUPPLY ASSEMBLY, Sheet 2 (For Non-Battery Units Only) UPC5100/UPC5110 Operation & Maintenance Manual...
Page 46
Figure 8-3. POWER SUPPLY ASSEMBLY (For Battery Units Only) Maintenance...
Page 47
DWG NO ANGLE 59037 MFG.ENG. CKD. THIRD ANGLE PROJECTION SV 8/25/87 RICE LAKE WEIGHING SYSTEMS APPROVED DES.ENG. JEW 8/25/87 DR.BY DJS 11/27/00 SCALE SHEET REVISION APPROVED Figure 8-4. CPU ASSEMBLY, Sheet 1 (For Battery Units Only) UPC5100/UPC5110 Operation & Maintenance Manual...
Page 48
Figure 8-5. CPU Assembly, sheet 2 (For Battery Units Only) Maintenance...
Page 49
Figure 8-6. CPU & POWER SUPPLY SCHEMATIC, Sheet 1 (For Non-Battery Units Only) UPC5100/UPC5110 Operation & Maintenance Manual...
Page 50
Figure 8-7. CPU & POWER SUPPLY SCHEMATIC, Sheet 2 (For Non-Battery Units Only) Maintenance...
Page 51
Figure 8-8. POWER SUPPLY SCHEMATIC, (For Battery Units Only) UPC5100/UPC5110 Operation & Maintenance Manual...
Page 52
Figure 8-9. CPU SCHEMATIC, Sheet 1 (For Battery Units Only) Maintenance...
Page 53
Figure 8-10. CPU SCHEMATIC, Sheet 2 (For Battery Units Only) UPC5100/UPC5110 Operation & Maintenance Manual...
UPC5100/UPC5110 Warranty and Return Policy If possible, please save original packing material which is specifically designed for the unit. Should it be necessary to ship the unit back to the factory, a suitable shipping container must be used along with sufficient packing material.
Need help?
Do you have a question about the UPC5100 and is the answer not in the manual?
Questions and answers