Ax60 Carbon Dioxide Detector Quick Start Guide - Quick Connect Option Analox Limited 15 Ellerbeck Court, Stokesley Business Park, North Yorkshire, TS9 5PT, UK T: +44 (0)1642 711400 F: +44 (0)1642 713900 W: www.analox.net E: info@analox.net...
Page 2
WARNING: READ THE SAFETY INFORMATION IN THIS QUICK START GUIDE BEFORE INSTALLING OR USING THE Ax60. WARNING: DO NOT TEST THE CO ALARM WHEN IT IS CLOSE TO THE EARS. IT HAS A HIGH VOLUME SOUNDER WITH A SOUND LEVEL OF 88 DECIBELS AT A DISTANCE OF 3 METRES.
Quick Start Guide - Quick Connect Option Introduction This Quick Start Guide explains how to install and use the Ax60 Quick Connect option. For more information on operation and maintenance, refer to the Ax60 User Manual P0159-800. For more information on servicing and calibration, refer to the Ax60 Service Manual P0159-803.
Page 5
Ax60 Carbon Dioxide Detector Quick Start Guide - Quick Connect Option Package contents 1 x Central Display, including: Ax60 Carbon Dioxide 1 x 2m factory fitted Quick Connect (QC) cable with white RJ45 Detector connector (for connection to CO Sensor)
Ax60 Carbon Dioxide Detector Quick Start Guide - Quick Connect Option Installation CAUTION: SOME ENCLOSURES ARE SUPPLIED UNFASTENED WITH FIXING SCREWS LOOSE. DO NOT OVER-TIGHTEN THE SCREWS WHEN FASTENING THE LIDS ON. Central Display Retain the clear protective film on the fascia until the installation is complete.
Ax60 Carbon Dioxide Detector Quick Start Guide - Quick Connect Option Components The Ax60 Quick Connect option is pre-wired with Cat5e cables and colour-coded RJ45 connectors for easy connection. The Quick Connect components are shown below. Central Display The Quick Connect Central Display is pre-fitted with two cable glands (see left).
Sensors or two CO Alarms on a common cable. Layouts In its simplest form a Quick Connect Ax60 system could incorporate a Central Display, one CO Sensor and one CO Alarm. A larger Ax60 system could incorporate a Central Display, four CO Sensors and eight CO Alarms.
Page 9
Ax60 Carbon Dioxide Detector Quick Start Guide - Quick Connect Option 1 x Central Display; 1 x CO Sensor; 2 x CO Alarms 1 x Central Display; 2 x CO Sensors; 4 x CO Alarms 1 x Central Display; 4 x CO Sensors;...
Ax60 Carbon Dioxide Detector Quick Start Guide - Quick Connect Option Connection The Central Display‘s terminal blocks are shown below. These must be connected to the power supply, beacon and relays (the Sensor terminal block has a cable already fitted).
Page 11
Ax60 Carbon Dioxide Detector Quick Start Guide - Quick Connect Option 5.1.1 PSU cable identification Plug-in type PSU cable identification Hard wired type PSU cable identification Black with stripe: Positive (24V) Black with stripe: Negative (0V) Black with print: Negative (0V)
Ax60 Carbon Dioxide Detector Quick Start Guide - Quick Connect Option Operation Central Display The Central Display is used to configure and operate the system. The three buttons on the front panel allow access to the software functions. The three LED (light-emitting diode) indicator lamps display the system status.
Page 13
Ax60 Carbon Dioxide Detector Quick Start Guide - Quick Connect Option Sensor The CO Sensor has a green Power indicator on the bottom left-hand part of the fascia. This is used to indicate the following conditions: Power indicator Under normal conditions the Power indicator flashes once per second to indicate that the power is on and the unit is operating.
Page 14
Quick Start Guide - Quick Connect Option Powering up 6.4.1 Power-on-self-test (POST) When you switch on the power supply the Ax60 performs a power-on-self-test (POST) which takes about 30 seconds. The results of the POST are displayed on the screen: Operator input Software response...
Page 15
Ax60 Carbon Dioxide Detector Quick Start Guide - Quick Connect Option Operator input Software response Central Display text Optional display text Wait for CO Sen- sors to warm up... This screen may display for a few seconds to show Sensor warm-up status.
( ) in each Sensor installed in the system. WARNING: DISCONNECT AND ISOLATE THE Ax60 SYSTEM FROM THE MAINS POWER SUPPLY BEFORE OPENING THE CO SENSOR ENCLOSURES. To access the jumper link, open the CO Sensor enclosure.