Intek Rheotherm 210 Manual

Precision flow meters designed to provide accurate representation of fluid flow rate.
Hide thumbs Also See for Rheotherm 210:

Advertisement

!
SECTION 1
GENERAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 1 -
1.1
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 1 -
1.2
DESCRIPTION OF OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 2 -
1.3
PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 2 -
!
SECTION 2
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 4 -
2.1
SENSOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 4 -
2.2
TRANSMITTER ELECTRONICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 6 -
2.3
ELECTRICAL CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 7 -
!
SECTION 3
OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 11 -
3.1
START UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 11 -
3.2
GENERAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 11 -
3.3
OPERATIONAL INTERFACES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 11 -
3.4
INSTRUMENT CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 12 -
3.5
OUTPUT CURVE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 16 -
!
SECTION 4
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 17 -
4.1
GENERAL MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 17 -
4.2
FLOW CALIBRATION ADJUSTMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . - 17 -
4.3
SPARE PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 17 -
4.4
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 17 -
!
SECTION 5
CUSTOMER SERVICE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 21 -
5.1
QUESTIONS ON EXISTING HARDWARE . . . . . . . . . . . . . . . . . . . . . . . . - 21 -
5.2
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 21 -
5.3
FACTORY AND FIELD SERVICE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 21 -
5.4
DECONTAMINATION OF EQUIPMENT . . . . . . . . . . . . . . . . . . . . . . . . . - 21 -
5.5
QUESTIONS ON NEW EQUIPMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 21 -
!
SECTION 6
CUSTOM INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 22 -
6.1
UNIT IDENTIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 22 -
6.2
CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 22 -
6.3
SPECIAL INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 22 -
TABLE OF ORIGINAL CALIBRATION DATA . . . . . . . . . . . . . . . . . . . . . . . . . . - 23 -
FLOW OUTPUT CURVE
MODEL 210 FLOW METER SENSOR INSTALLATION
TABLE OF CONTENTS
Intek, Inc. 2010
M an ua l no . A2 10 06 06 Re v. B
MODEL 210FM RevB.wpd

Advertisement

Table of Contents
loading

Summary of Contents for Intek Rheotherm 210

  • Page 1 TABLE OF ORIGINAL CALIBRATION DATA ......- 23 - FLOW OUTPUT CURVE MODEL 210 FLOW METER SENSOR INSTALLATION Intek, Inc. 2010 M an ua l no . A2 10 06 06 Re v. B MODEL 210FM RevB.wpd...
  • Page 2 Intek's part. The purchaser will assume all responsibility and expense for removal, decontamination and reinstallation of equipment. Rheotherm flow meters are manufactured under United States patent numbers 4,255,968; 4,942,763; 4,949,578; 5,485,754; 5,752,411 and 6,526,755. Intek, Rheotherm, RheoVac, Rheovec, Rheomax and RheoSmart are registered trademarks of Intek, Inc. Intek, Inc.
  • Page 3: Section 1 ! General Information

    Rheotherm instruments are manufactured exclusively by Intek, Inc. and employ a patented thermal technique used by industry since 1978. The unique sensor designs have protected sensors, are easy to install and require little or no maintenance.
  • Page 4: Description Of Operation

    The rate at which heat is removed from the heated sensor by the stream is related to fluid velocity. Hence, the measured temperature differential between the reference sensor and heated sensor is a function of flow rate. Intek, Inc. is licensed to use this patented and trademarked flow measurement method.
  • Page 5 Any special information pertaining to your unit is covered under CUSTOM INFORMATION (SECTION 6). Time should be taken to carefully read these instructions prior to installation of the equipment. Should any questions arise or problems occur, call Intek for immediate assistance.
  • Page 6: Section 2 ! Installation

    SECTION 2 INSTALLATION 2.1 SENSOR !! WARNING: If the instrument has been selected for use in hazardous environments the model number shall end with the suffix -FM. The -FM notation designates that the instrument has been designed, manufactured for use, and reviewed by Factory Mutual (FM) as suitable for use in Class I, Division 1, Groups B, C, D, (Class I, Zone 1, IIB+H2 for US Only), Class II, Division 1, Groups E, F, G, Class III, Division 1, type NEMA 4X, T6 Ta=60...
  • Page 7 the line is via compression fittings, hose with clamp, threaded fittings or flanges, whichever is appropriate. Care must be taken not to transmit a twisting force through the sensor’s midsection. The TU and TUL sensors, whether flanged or not, must not be used to pull other piping together or to make up angular mismatch of fittings.
  • Page 8: Transmitter Electronics

    be avoided. Sudden pressure drops can cause the absorbed gas to release into the liquid, making the flow sporadic and difficult to measure. Control valves should be placed down stream of the sensor. The ideal installation will provide the sensor with well established smooth flow, uniform system temperature and consistent fluid media.
  • Page 9: Electrical Connections

    !! WARNING: When used in hazardous environments, the model number will include the -FM suffix. Using the shaft of a screwdriver placed between the lid’s center bosses, the sensor mounted box lid shall be tightened enough to assure it cannot be removed without the use of a tool.
  • Page 10 type has been ordered it will be listed in SECTION 6.3 - SPECIAL INSTRUCTIONS. (Active: current to the loop is sourced by transmitter. Passive: output receiver sources the current.) Pull wires through the conduit. Wire for the power connection must be no smaller than 22 gauge or as required by applicable local or company wiring codes.
  • Page 11 Figure 1 - Electrical Interface Model 210R option - 9 -...
  • Page 12 - 10 -...
  • Page 13: Section 3 ! Operation

    100%. During startup of units with a display, the alphanumeric display will show ‘*INTEK, INC.*’ on the top line and ‘(614)895-0301’ on the bottom display line. After fifteen to sixty seconds (depending on flow meter response) the reading will stabilize.
  • Page 14: Instrument Configuration

    Information Section). The instrument output will extend outside of the 4-20 mA range to signify an alarm condition. A low value will indicate a problem has been detected with the sensor. All other error types will produce an output value higher than 20 mA. The only expected time the signal will be outside the 4-20 mA level is for a few seconds after powering on the instrument.
  • Page 15 A quick reference guide is provided in the form of Table I which contains a list of available con- figuration options accessible using the display and the proximity sensors mentioned in SECTION 3.3 Local Display Configuration The Model 210 instrument with display option offers a user interface comprised of a 2x16 character LCD display, a status LED, and two proximity sensors.
  • Page 16 Table II. Display Parameter List Mo del 210 Display Parameter List Name Parameter Description Flow Flow Rate Rate of flow in customer specified units Temp °C Fluid Temperature Fluid temperature in degrees Celsius Temp °F Fluid Temperature Fluid temperature in degrees Fahrenheit Total Flow Total accumulated flow volume or mass Serial Number...
  • Page 17 For Non-Hazardous Locations: NOT ALL MODEL 210 INSTRUMENTS ARE !! WARNING: FOR USE IN HAZARDOUS LOCATIONS. ONLY INSTRUMENTS WITH A ‘-FM’ DESIGNATION IN THE MODEL NUMBER ARE APPROVED FOR USE IN HAZARDOUS AREAS. To perform a calibration adjustment on an instrument without a display, the transmitter housing cover must be removed to access the IR communications port and a palm device is required.
  • Page 18: Output Curve

    A message of ‘Flow is too low / for Adj High’ or ‘Flow is too high / for Adj Low’ will appear during calibration if the flow is out of the allowed range for the adjustment. Adjust Calibration Using Palm Device Refer to Palm Software Appendix. Restore Factory Calibration Restore the factory calibration settings by setting both the top and bottom display lines to the ‘Software Version’...
  • Page 19: Section 4 ! Maintenance

    SECTION 4 MAINTENANCE 4.1 GENERAL MAINTENANCE Certain precautions should be taken to ensure proper performance of all models of flow instruments. Since the measurement technique involves a signal resulting from heat transfer to the flowing medium, care should be exercised to prevent build-up of varying layers on the walls of the sensor. Layers such as bacterial growth, dried paints, gas bubbles and non-solubles can result in measurement below actual flow rates.
  • Page 20 ‘Customer Mode’ when finished with diagnostics.) Viewing Instrument Status Using Palm Device Activate the Intek Palm User Interface software (IntekPUI) on a IR equipped palm device. Tap the black ‘IntekPUI’ menu bar to reveal the options menu and select the ‘Troubleshooting’...
  • Page 21 TABLE IV. Troubleshooting Guide - Instrument Diagnosed Problems INDICATION PROBABL E CAUSE ACTION STATUS CODE 1. Nominal operation 1. N/A ‘00’ 1. Check contacts and continuity of STATUS CODE 1. Cable cut or all sensor connections cable and check sensor for signs ‘10’...
  • Page 22 covers removed are not recommended. If you are performing any diagnostic operation in a hazardous environment proceed with caution and abide by all applicable safety guidelines. Although only low voltages (<24Vdc) are present, limited energy sparking may still occur in the event of a short while making these measurements. When completed, all enclosure covers should be replaced and properly fastened per instructions of SECTION 2.3.
  • Page 23: Section 5 ! Customer Service

    Rheotherm instrument you also receive Intek's outstanding customer service. For sales or product service, call your local representative or Intek directly at (614) 895-0301, 8AM to 5PM EST/EDT weekdays or fax us anytime at (614) 895-0319. E-mail inquiries should be sent to sales@Intekflow.com or techsupport@Intekflow.com.
  • Page 24: Section 6 ! Custom Information

    SECTION 6 CUSTOM INFORMATION UNIT IDENTIFICATION Model no.: Serial no.: Customer identification: CONFIGURATION The configuration of this unit, as originally shipped from the factory: Input Power: Q 24 Vdc Q Other Q Display Q Hazardous Location: Class I, Division 1, Groups B, Options: C, D, (Class I, Zone 1, IIB+H2 for US Only), Class II, Division 1, Groups E, F, G, Class III, Division 1, type...
  • Page 25: Table Of Original Calibration Data

    TABLE OF ORIGINAL CALIBRATION DATA FOR FUNCTIONAL TEST Unit Serial Number - Instrument Flow Output Flow Rate TEMP Display Note: An offset of data in column III (with respect to column II) may appear if the instrument has been field adjusted. - 23 -...
  • Page 26 Model 210 Rheotherm ® Flow Instrument Palm Software Appendix...
  • Page 28: Table Of Contents

    PALM DEVICE PROBLEMS ........- A7 - ©Intek, Inc. 2005 Manual no.
  • Page 30: Palm Device General Information

    SECTION 1 PALM DEVICE GENERAL INFORMATION The Rheotherm ® Model 210 flow instrument has the ability to communicate with a standard off-the- shelf palm device. This allows an operator to configure, calibrate, and access instrument status in a safe, quick, and easy manner. Using a palm device with custom Model 210 software, a wireless infrared (IR) link is established without the use of additional tools or hardware.
  • Page 31 !! !! WARNING: To perform a calibration adjustment for a "blind" instrument, the transmitter housing cover must be removed to access the IR communications port and a palm device is required. Since the instrument must be active to perform this operation, follow prudent safety procedures before attempting this procedure.
  • Page 32: Operation

    IR port. After installing IntekPUI onto the palm device, locate the application and tap on the Intek logo to launch the application. The ‘Flow Calibration Adjustment’ option will auto-launch and the palm will begin to initialize communication with the Model 210.
  • Page 33: Calibration Adjustment

    First the flow must be steady and the desired flow rate indication must already be determined at the low flow rate. Activate the Intek Palm User Interface software (IntekPUI) on a IR equipped palm device. The ‘Calibration Adjustment’ option will open automatically.
  • Page 34: Calibration Selection

    low and high values are at least 10% of full scale flow apart. E.g. if the full scale flow is 10 cc/min and your normal operating range is 6.5 - 7 cc/min then perform a low adjustment at 6 cc/min and a high adjustment at 7 cc/min. Restoring the Factory Calibration At any time the original factory calibration can be restored.
  • Page 35 Figure 6 - Display Configuration Form Change Display Navigate to the ‘Display Configuration’ option form. Holding the palm device 1 to 2 feet in front of the instrument window, select the – desired display line (Top Display or Bottom Display) by tapping the < >.
  • Page 36: Troubleshooting

    Troubleshooting with the assistance of a palm device is covered below. VIEWING INSTRUMENT STATUS USING PALM DEVICE Activate the Intek Palm User Interface software (IntekPUI) on a IR equipped palm device. Tap the black ‘IntekPUI’ menu bar to reveal the options menu and select the ‘Troubleshooting’...
  • Page 37 Locate the Intek supplied ‘IntekPUI.prc’ source file and reinstall the application per your palm device’s install tool directions. Make sure to perform a ‘HotSync’ operation per palm device’s instruction to complete the installation. No Communication Navigate to the ‘Calibration Adjustment’ option form and hold the palm device directly in front of the Model 210's IR port.
  • Page 38 Model 210 ® Rheotherm Flow Instrument User Interface Software (UIS) ® for Windows based PCs Appendix...
  • Page 40 INSTRUMENT CALIBRATION ....... . . - B5 - ©Intek, Inc. 2007 Manual no.
  • Page 41: General Information

    !! !! WARNING: If the instrument is installed in a hazardous area do not open the instruments covering to perform live adjustments. Check your PC’s hazardous area rating; ensure compatibility before using it in connection with the Rheotherm flow instrument. IR PORT INTEK SCROLL SELECT Figure 1 - Instruments with “Display”...
  • Page 42: Installation

    SECTION 2 INSTALLATION !! CAUTION: The UIS is designed for instrumentation configuration and maintenance, and not required under normal operating conditions. The base model 210 has successfully completed all EMC tests as required for CE declaration. However the flow instrument, communications path, and the PC should be evaluated for CE compliance, where applicable, by the system integrator.
  • Page 43: Uis Installation

    3. UIS INSTALLATION a. PC Recommendations for best performance: i. Windows XP or later. ii. Pentium equivalent or better. iii. 256 MB of RAM or greater. iv. XVGA video mode 800x600 to 1088x612. b. Locate the Rheotherm UIS installation file (typically supplied on CD). Run the setup.exe or follow the auto-run prompt to install the application.
  • Page 44: Operation

    SECTION 3 OPERATION !! CAUTION: Adjustments to the calibration will override the factory calibration settings. Before field calibrating the unit, make sure that the suspected indication errors are not due to installation or other factors by reviewing the installation guidelines and making any necessary system changes.
  • Page 45: Instrument Calibration

    b. Local Display Configuration (Applicable only to -IDT or -XDT models) i. Click on the [Config][Display] tab to reveal the display configuration form (Figure 6). ii. The local LCD display can provide any item in the gray top and bottom pick lists. Choose the desired parameter and press the corresponding yellow “Disp”...
  • Page 46 Note: If making an adjustment when the instrument is indicating zero flow it is important to assure that there is actually flow passing through the sensor. A zero flow indication is possible when flow is above the calibration’s low limit. In this case, consecutive flow adjustments may be needed to stabilize the flow indication as the instrument adjusts itself away from the shutdown control limit.
  • Page 47 Figure 9 - Calibration Select Form Figure 10 - Calibration Restore Form ii. Change the ‘Select calibration’ pull-down list to load the desired calibration channel. Examine the ‘Active Calibration’ indicator for confirmation of the change. c. Calibration Restore i. Restores the active calibration channel to the factory set values (Figure 10). d.
  • Page 48 Model 210 ® Rheotherm Flow Instrument Modbus Communications Appendix...
  • Page 50 READ HOLDING/INPUT REGISTERS – 03 (0x03 / 04 (0x04)……………. -C5- CHANGE SLAVE ADDRESS – 10 (0x0A)…………………………………. -C6- CHANGE BAUD RATE – 13 (0x0D)……………………………………….. -C7- CHANGE CALIBRATION – 14 (0x0E)…………………………………….. -C8- Intek, Inc. 2010 Manual no. A210 MBus 0810 Rev. – MODEL 210 MODBUS Appendix.doc...
  • Page 52: Section 1 - Model 210 Modbus Communication

    SECTION 1 - MODEL 210 MODBUS COMMUINICATION 1. OVERVIEW This appendix describes the implementation of the standard Modbus protocol, which is used to retrieve process variables for the Model 210 flow instrument. If Modbus mode is enabled the IR buttons on the Model 210 and the serial client user interface do not work. 2.
  • Page 53: Modbus Order

    Diagnostics Get Event Counter Get Event Log Write Multiple Registers Report Slave ID The following is a list of the Intek specific Modbus functions the Model 210 supports. Supported Custom Modbus Functions Function Hex Code Change Slave Address Change Baud Rate...
  • Page 54: Register Definitions

    6. REGISTER DEFINITIONS The Model 210 conserves memory space by not supporting coils or discrete inputs. This creates more space for holding and input registers where the process variables are stored. This modification does not affect the formation of a Modbus request packet, which will be demonstrated in the packet section.
  • Page 55: Packet Definitions

    7. PACKET DEFINITIONS Cyclic Redundancy Check (CRC) The 2 Byte CRC is required to be appended to the end of every Modbus packet and is implied in the following references. Exception Packets If an error is detected in one of the following packet definitions a certain error code is applied and sent back in an exception response.
  • Page 56: Read Holding/Input Registers - 03 (0X03 / 04 (0X04)

    8. READ HOLDING/INPUT REGISTERS – 03 (0x03) / 04 (0x04) These are the contiguous block of registers referenced above in the Register Definitions section. The Model 210 stores the process variables as two 8 bit hex values therefore you must read the registers as multiples of two.
  • Page 57: Change Slave Address - 10 (0X0A)

    CHANGE SLAVE ADDRESS – 10 (0x0A) This command changes the slave address of the desired Model 210 instrument. This change is made immediately and does not require a restart. Request Type Example Device Address 1 Byte (1 - 247) 0x01 Function code 1 Byte 0x0A...
  • Page 58: Change Baud Rate - 13 (0X0D)

    10. CHANGE BAUD RATE – 13 (0x0D) This command switches the baud rate to one of the other two supported rates. The baud rate is changed immediately, however, if a display is being used then a power cycle is required to initialize the new baud rate.
  • Page 59: Change Calibration - 14 (0X0E)

    11. CHANGE CALIBRATION – 14 (0x0E) This command switches the current calibration to one of the other 3 available calibration channels. The new calibration value should be a hex value between A-D (10-13 integer) corresponding to the desired letter of calibration. Request Type Example...

This manual is also suitable for:

Rheotherm 210 fm

Table of Contents