Mantracourt LCI User Manual

Load cell junction box with fault monitor

Advertisement

Quick Links

LCI User Manual
MICRON METERS
www.micronmeters.com
metersinfo@micronmeters.com
LCI
Load Cell Junction Box with Fault Monitor

Advertisement

loading

Summary of Contents for Mantracourt LCI

  • Page 1 LCI User Manual MICRON METERS www.micronmeters.com metersinfo@micronmeters.com Load Cell Junction Box with Fault Monitor...
  • Page 2: Table Of Contents

    Contents Chapter 1 Introduction to the LCI ...................... 2 Chapter 2 Installing the LCI ......................3 Chapter 3 Setting up the LCI ......................4 Sequence of Operations ......................... 4 Chapter 4 Specifications of the LCI ..................... 7 Wiring Conditions ......................... 8 Warranty............................
  • Page 3: Chapter 1 Introduction To The Lci

    4V and 12V without requiring any special set-up procedure or link options. Host requirements: The Host must be capable of supplying power to the LCI as well as excitation for the load cells (see Chapter 4 for specifications).
  • Page 4: Chapter 2 Installing The Lci

    LCI Connection Details Important Note: When connecting less than four load cells to the LCI, start at channel one and fill the remaining channels in numerical order. Set the ‘no C’ parameter accordingly (see table below). LCI Connections to LCB & ADW15 Details...
  • Page 5: Chapter 3 Setting Up The Lci

    Chapter 3 Setting up the LCI The LCI features a 4-digit red LED display, four programming keys and two function keys. On ‘Power Up’, the display will show either ‘good’ or display an error code e.g. ‘1Er5’ which, in this case, would refer to Load Cell 1 with error condition 5.
  • Page 6 This is the time (in seconds) an error must be present before the error alarm is displayed and output via the relay. The Error Displays on the LCI Note: On Power Up the display will show either ‘good’ or (‘...
  • Page 7 The alarm relay is energised during a healthy state and drops out if an error occurs (fail safe). The Millivolt Display readings on the LCI To view the average of all of the load cell signals, press the ‘mV’ button.
  • Page 8: Chapter 4 Specifications Of The Lci

    Connections 2 part terminals, up to 2.5mm² cable 4 x 5 way for load cell connection 1 x 5 way for LCI output and power input 1 x 3 way alarm relay contacts Dimensions 200 x 120 x 75mm. (PCB dimensions 170 x 100mm excluding mounting...
  • Page 9: Wiring Conditions

    (3) years from the date of dispatch. If the 'Mantracourt' product you purchase appears to have a defect in material or workmanship or fails during normal use within the period, please contact your Distributor, who will assist you in resolving the problem.
  • Page 10 LCI User Manual Applies To: Part Number: 517-164 Issue Number: Dated: 22 November 2016 In the interests of continued product development, Mantracourt Electronics Limited reserves the right to alter product specifications without prior notice. www.mantracourt.com Mantracourt Electronics Limited LC1 User Manual...