Command
Value
Function:
10h
Data:
0000h to FFFFh
0001h to 007Bh
2 x N
Response
Value
Function:
10h
Data:
0000h to FFFFh
0001h to 007Bh
A parameter value is stored in a 16-bit register and always requires, therefore, a
2-byte length. The most significant byte precedes the least significant one.
Table 8.C (page 46) lists all available parameters with their register number.
The write function 10h changes only the value in the RAM memory; changes
are therefore deleted in case of power supply failure in the instrument. To
make any parameter changes permanent, "Parameter permanent storage" command
shall follow the parameter change (function 05h).
To change parameters protected by an access code, you must first write the code in
the configuration registers from number 10000 to 10005. After changing the
parameters, modify the content of the above mentioned registers to disable the code.
If the request involves entering a parameter not included in Table 8.C, the instrument
replies with the following error message:
Response
Value
Function:
90h
Data:
02h
Example:
Suppose you want to change CO
values:
0V or 4mA = 350 ppm
10V or 20mA = 1200 ppm.
The two parameters are protected by the access code "240471".
In order to change the parameters, the access code has to be transmitted first by
means of the command below:
HD45 / HD46
Length
Description
1 byte
Parameters writing command
2 bytes
First parameter address
2 bytes
Number of parameters to be changed (N)
1 byte
Number of bytes
2 x N bytes
New parameters values
Length
Description
1 byte
Parameters writing command
2 bytes
First parameter address
2 bytes
Number of changed parameters (N)
Length
Description
1 byte
Error in parameters writing
Operation involved a non-existent
1 byte
parameter (Tab. 8.F page 59)
values corresponding to the initial and full scale
2
- 55 -
V2.4