Control Solution Testing

You should do a control solution test when you are not sure of your test
strip results and want to check that your Reader's built-in meter and test
strips are working properly .
IMPORTANT:
Control solution results should fall within the control solution range
printed on the test strip instructions for use .
Do NOT use control solution past the expiration date . Discard
control solution 3 months after opening or on the expiration date
printed on the bottle, whichever comes first . (Example: open April
15, discard July 15; write the discard date on the side of the bottle .) .
The control solution range is a target range for control solution
only, not for your blood glucose results .
The control solution test does not reflect your blood glucose level .
Use only MediSense (low, medium or high) Glucose and Ketone
Control Solution with the Reader's built-in meter .
Check that the LOT number printed on the test strip foil packet and
instructions for use match .
Replace the cap securely on the bottle immediately after use .
Do NOT add water or other liquid to the control solution .
Contact Customer Service (1-855-632-8658) for information on how
to obtain control solution .
71
ART38553-101_rev-A_manual.indd 71-72
Step
Action
1
From the Home Screen, touch the Settings
10:23pm
Ends in 10 days
. Touch Control Solution Test to do
symbol
a control solution test .
Check
Review
Glucose
History
Settings
Sounds
Target Range
Control Solution Test
1 / 3
2
Check the test strip expiration date .
3
Open the foil test strip packet at the notch and
tear down to remove the test strip .
72
2/1/18 2:13 PM