Contents PREFACE INTENDED USE GENERAL DESCRIPTION OF MAJOR COMPONENTS WARRANTY INFORMATION SPECIFICATIONS TROUBLESHOOTING SERVICE INSTRUCTIONS WIRING DIAGRAM SPARE PARTS LIST DRUCKER DIAGNOSTICS SM029 HORIZON 24 FLEX SERVICE MANUAL REV A...
Failure to follow the procedures in this document may result in personal injury or instrument damage. Drucker Diagnostics will not be held liable for any injury or damage because of improper servicing.
Tighten rotor screw per section 7-2 PROBLEM POSSIBLE CAUSE SOLUTION Debris in air intake / exhaust ports Remove power before clearing debris. Whistling noise while running Gasket failure Requires service DRUCKER DIAGNOSTICS SM029 HORIZON 24 FLEX SERVICE MANUAL REV A...
Rotor damaged Replacement required ‘ROTOR’ Software selection of rotor is unknown Service Required – Call Drucker technical service. 7. SERVICE INSTRUCTIONS 7.1. Cleaning • The cabinet, rotor and accessories can be cleaned using soap and water, isopropyl alcohol, or a 10% (5500 PPM) bleach solution •...
Page 6
– do not over tighten. Each nut sets the height of a separate control panel button. Too low, and they cannot be actuated, too high and they jam against DRUCKER DIAGNOSTICS SM029 HORIZON 24 FLEX SERVICE MANUAL REV A...
Page 7
Attach the motor to the motor mount plate with four #8 screws and torque to 18 in-lbs (Fig 2). • Secure the motor wire harness with a zip tie as shown. Snip the tag end of the zip tie flush when complete. DRUCKER DIAGNOSTICS SM029 HORIZON 24 FLEX SERVICE MANUAL REV A...
Page 8
(fig 7). • Install one fender washer onto each stud (fig 8). • Finish the assembly with one Nylok nut on each stud, and tighten to 13 in-lbs DRUCKER DIAGNOSTICS SM029 HORIZON 24 FLEX SERVICE MANUAL REV A...
Page 9
Secure the base with six #8 screws torqued to 10 in-lbs. (fig 11). • Install the exhaust cover and secure with three #8 screws torqued to 10 in-lbs. (fig 12). DRUCKER DIAGNOSTICS SM029 HORIZON 24 FLEX SERVICE MANUAL REV A...
Page 10
Carefully place the cabinet onto the base taking care not to pinch any wires between the two. • Complete the assembly by replacing the eight #8 screws, with a 3/32 hex key. DRUCKER DIAGNOSTICS SM029 HORIZON 24 FLEX SERVICE MANUAL REV A...
Need help?
Do you have a question about the Horizon 24 and is the answer not in the manual?