Calibration tester

Advertisement

Available languages

Available languages

Directions for Use
Welch Allyn SureTemp and SureTemp Plus
Thermometers
Caution Store thermometers for testing in the same
room as the 9600 Plus Calibration Tester for
approximately 30 minutes prior to testing to allow the
thermometer to reach ambient room temperature.
Note
Prior to testing visually inspect the probe for any physical
damage. Damaged probes must be replaced.
Figure 4. 9600 Plus Calibration Tester with the Welch Allyn SureTemp
Plus Thermometer
To begin calibration verification of the thermometers:
1. Remove the probe from the probe well and clean it with either
a 70% isopropyl alcohol solution, a 10% chlorine bleach
Chapter 3 Operation
13