PROCEDURE FOR HANDHELD PEN FLUOROMETER

Initial Setup Procedure

1.
Press "SELECT" button on fluorometer. "PF" will blink on screen.
2.
Press "READ/ENT". Screen will show 100.0.
3.
Press "SELECT" button twice and the first 0 will change to 1 and then 2.
Screen will show 200.0. This will set "PF" to 200.
4.
Press "READ/ENT" 4-5 times till screen shows trA. Screen should show
"000.0". Press "READ/ENT" 7-8 times till screen goes blank. trA is now set
to 0.
5.
Pen fluorometer is now ready for calibration/testing.

Calibration Procedure

1.
Make sure clear rubber plug is inserted in bottom of pen fluorometer, hold
pen horizontally.
2.
Rinse and dump the well three times with Distilled Water.
3.
After dumping the third rinse, fill the well with Distilled Water.
4.
Press the "CAL" button.
5.
The screen will read "2-Pt CAL." After a few seconds, the screen will
change and read "dIP H2O."
6.
Press "READ/ENT".
7.
Screen will blink "MEAS" several times (for about 10-20 seconds) and then
will read "dIP Std".

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TRASAR 2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Nalco TRASAR 2

  • Page 1 PROCEDURE FOR HANDHELD PEN FLUOROMETER Initial Setup Procedure Press “SELECT” button on fluorometer. “PF” will blink on screen. Press “READ/ENT”. Screen will show 100.0. Press “SELECT” button twice and the first 0 will change to 1 and then 2. Screen will show 200.0. This will set “PF” to 200. Press “READ/ENT”...
  • Page 2 NAMPAC May 1, 2008 Page 2 Rinse and dump the well three times with TRASAR Calibration Standard S0920. After dumping the third rinse, fill the well with S0920 calibration standard. 10. Press the “READ/ENT” button 11. The screen will blink “MEAS” several times (for about 10-20 seconds). 12.

Table of Contents

Save PDF