Open the Signal Conditioner Unit
Disconnect all signal input and outputs before attempting this procedure.
1) To open the Signal Conditioner Unit—Remove Side Covers from Unit by removing 4 screws, using
Instructions
a Phillips head screwdriver.
2) Short TP6 & TP17 together, either by wire or, if available, jumper S1-A.
3) After shorting TP6 to TP17, apply power to the unit.
4) The Front Panel Power Indicator (LED) will NOT turn on.
5) After power has been applied to the unit for about 3 seconds, remove jumper S1-A or open the TP6
& TP17 connection without turning off the power.
Solder wire from
TP6 to TP17 on
solder side of board.
Version A
6) The communication parameters now have been restored to the default settings temporarily in
RAM memory. (If power is removed from the unit, the RAM memory will be lost and the
communication parameters will be lost.)