Switching on the device ..................28 Gateway GC 9310 ..................28 4.2.1 GC 9310 commissioning ..................30 4.2.2 Interfaces and logs ....................32 4.2.3 Data storage ......................33 4.2.4 Electrical connections of the GC 9310 ..............34 Manual GC 9300 EN14 July, 24th 2018...
Page 4
5.1.8 User: User-specific display .................. 45 5.1.9 Detail: List of all measurements and parameters of the GC 9300 ....... 46 5.1.10 Archive: Display of archive entries ..............47 5.1.11 Log: Display of the log books ................48 5.1.12 DSfG - Archive: Display of the DSfG archive ............49 5.1.13 DSfG - Delete: Deletion of the DSfG archive ............
Howev- er, its use can entail dangers that are avoidable by complying with this manual. The analytical computer GC 9300 must only be used as intended and in technical- ly sound condition.
This notice can provide you with helpful tips to make your work easier. This notice also provides you with further information about the device or the work process in order to prevent operator error. Manual GC 9300 EN14 July, 24th 2018...
Note All notices in the manual must be observed. Use of the analytical computer GC 9300 is only permitted in accordance with the specifications in the operating manual. RMG assumes no liability for damages arising due to disregard of the oper- ating manual.
Page 8
The technical specifications must be observed and followed for safe opera- tion (Chapter 6 Technical data). Performance limits must not be exceeded. For safe operation, the analytical computer GC 9300 must only be used in the scope of the intended use (Chapter 1.3 Function to 1.5 Area of applicabil- ity) The GC 9300 complies with current standards and regulations.
Page 9
With the danger of explosion, the following must be observed, in particular: The analytical computer GC 9300 is not approved or designed for use in ex- plosion-prone areas. Installation must only take place in safe rooms. The analytical computer is intended for installation in a control cabinet in an electronics room.
Page 10
1 About this manual Danger In general, the replacement of an analytical computer GC 9300 must only • be carried out by RMG Service. 1.1.4.3 Dangers during maintenance and repair Operating personnel The operating personnel use and operate the device in the scope of the intended use.
Risk assessment and minimization According to assessment by qualified employees of RMG, the analytical computer GC 9300 is subject to risks during its use. Risks can arise, for instance, during use outside of the permissible temperature range. Impermissible current and voltage values can trigger explosions in explosion-prone areas.
Applicability of the manual This manual describes the analytical computer GC 9300. The analytical computer GC 9300 is only part of a complete system. The manuals of the other components of the system must be observed. If you find contradictory instructions, contact RMG and/or the manufacturers of the other components.
1 About this manual 1.1.6.2 Dangers of operation in EX areas The analytical computer GC 9300 is not intended for use in explosion-prone areas. Only operate the analytical computer GC 9300 in fault-free and complete condi- tion. If you make technical changes to the device, safe operation can no longer be guaranteed.
1.1.10 Storage Avoid extended periods of storage. Inspect the analytical computer GC 9300 for damage and correct function after storage. Contact the RMG service department to arrange for inspection of the device after a storage period of longer than one year.
GC 9300. The second part de- scribes how the GC 9300 works and its intended purpose. The second chapter explains the basic function and operation of the GC 9300 and data storage – taking place in accordance with the DSFG standard, in particular.
1 About this manual 1.3 Function The analytical computer GC 9300 is a component of the process gas chromato- graph PGC 9300, which analyzes natural gas and biogas and provides the neces- sary measurement values to determine the energy content and calculate the com- pressionability factor.
ISO 6976 (except for the methane number). 1.5 Area of applicability The analytical computer GC 9300 is an essential element of the overall measuring unit, which analyzes natural gas and biogas and provides the necessary meas- urement values to determine the energy content and calculate the compression ability factor.
For connection of USB components (e.g. a mouse), sealed in custody transfer mode. HOME button To switch between the start screen and error screen. Touch screen Display and control panel of the analytical computer. Manual GC 9300 EN14 July, 24th 2018...
Further operating possibilities are available via DSfG bus with RMG software products such as the PGC revision program AKA-II. 2.2 Operation Operation takes place on the analytical computer GC 9300 and enables the follow- ing actions: • Reading of analysis results •...
Page 20
2 Introduction • NORMAL CALIB. Manually started calibration (equivalent to automatic cal.) • REF GAS Test gas analysis (reference gas) Manual GC 9300 EN14 July, 24th 2018...
Retention times (20) • Retention times with the basic calibration • Response factors (20) • Response factors with the basic calibration (20) • Peak start time (20) • Peak end time (20) • Total area Manual GC 9300 EN14 July, 24th 2018...
Analog values 1 to 16, status bit string 2280 Extended Hu, Wo, Wu, MZ, Zn, 2280 Hourly averages unnormalized total components Reference gas Ho, rhon, CO , bit string Log book Event, Ho, rhon, components 2280 Manual GC 9300 EN14 July, 24th 2018...
3 Installation Installation 3.1 Installation The analytical computer GC 9300 is intended for installation in a control cabinet in a non-Ex area. Installation takes place in a dedicated rack; the device dimensions are: W x H x D = 213 x 128.4 x 310 mm (42 TE / 3 HE) Danger The GC 9300 is not intended for use in explosion-prone areas.
Collective message for digital outputs 9 - 12 Digital output 8 - Collective message for digital outputs 9 - 12 Warning contact + PGC: Warning Contact open in the event of a warning Warning contact - PGC: Warning Manual GC 9300 EN14 July, 24th 2018...
Page 25
(active + / passive -) Temperature input 1 supply ++ Room temperature Temperature input 1 sense + Room temperature Temperature input 1 sense - Room temperature Temperature input 1 supply -- Room temperature Manual GC 9300 EN14 July, 24th 2018...
Not allocated 3.2.2 Power supply 230 V/AC version 24 V/DC version 100 – 230 V Not allocated Not allocated + 24 V 100 – 230 V - 24 V Potential equalization Potential equalization Manual GC 9300 EN14 July, 24th 2018...
X12: COM 2 RS 232 WAGO-IO / GPS 170 RS 232 / RS 485* X13: COM 3 DSfG / Modbus RTU / Modbus ASCII / RMG-Bus X14: COM 4 RS 232 / RS 485* DSfG / RMG-Bus X15: COM 5...
Page 29
• 750-602 24 V DC power supply (galvanically isolated potential feed) The next figure shows the above setup in an enlargement of the previous setup; the numbers of the Wago modules can be read. Manual GC 9300 EN14 July, 24th 2018...
Page 30
Wago modules to the 02 COM 2 interface of the GC9300 family. To use the Wago modules, the following parameters on the GC93XX have to be set in the menu item 16 serial ports, 02 COM 2 as shown in the next figure. Manual GC 9300 EN14 July, 24th 2018...
Page 31
2.003 and higher. It may therefore no longer match that of existing installa- tions. Therefore, firmware updates should be avoided or should be real- ized from the RMG's service. Manual GC 9300 EN14...
The gateway GC 9310 is optional and designed to extend the 2 DSFG interfaces of the GC 9300 to 4 DSFG interfaces (GC 9300 + GC 9310); if more than two additional DSfG buses are required, additional GC 9310 units can be connected.
Page 33
RMGViewGC. Configuring GC 9300 for operation with GC 9310 In order to ensure that the GC 9300 Controller can work in combination with one or several GC 9310 gateways, the following settings must be made: •...
For this purpose, use of the same time sync source is recommended for the two devices. Like the GC 9300, the gateway GC 9310 is intended for installation in a control cabinet in the electronics room. Installation takes place in a dedicated rack; the device dimensions are: W x H x D = 213 x 128.4 x 310 mm (42 DU / 3 HU)
Page 35
Like the GC 9300, the individual screens are arranged like tabs: The values that the GC 9310 has received from the GC 9300 are shown in the table. All operating parameters, the number of the measuring element and the time stamp of the analyses are shown.
ISO 6976, and the optional calculation of the methane number. Like the GC 9300, entry of the code number and/or opening of the custody transfer switch is normally necessary for nearly all actions other than the display functions.
The data storage of analysis results is possible in non-custody-transfer measure- ment archives and in custody transfer archives in accordance with the DSFG standard; storage is no different than in the GC 9300 (chapter 2.3 Data storage) Manual GC 9300...
Not available Not available Not available Not available Not available Not available Not available Alarm contact + PGC: Alarm + Contact open in the event of an alarm Alarm contact - PGC: Alarm - Manual GC 9300 EN14 July, 24th 2018...
• Error status with color-coding of the “Error” tab • Ready indicator o “READY” for “Measuring element is ready” o “NOT READY” for “Measuring element is not ready” o “INITIALIZING” for “Measuring element starting up” Manual GC 9300 EN14 July, 24th 2018...
Page 44
Status displays, measuring element and digital IOs Service Service functions User User-specific display Detail List of all measurements and parameters of the GC 9300 Archive Display of the GC archive Display of the log books DSfG Display of the DSfG archive...
Stream-2 Measuring gas from Gas Stream 2 Stream-3 Measuring gas from Gas Stream 3 Stream-4 Measuring gas from Gas Stream 4 REF gas Last reference gas analysis Cal. gas Last calibration gas analysis Manual GC 9300 EN14 July, 24th 2018...
There is a “Reject” selection point here for SKIP analyses for the purpose the flushing and the corresponding results were not as- signed to a gas stream. Manual GC 9300 EN14 July, 24th 2018...
Variables such as pressure and temperature for Columns 1 to 3 and inlet pres- sures for measuring gas and carrier gas can be found here for the entries in the maintenance log. An additional, second carrier gas is used for some applications. Manual GC 9300 EN14 July, 24th 2018...
Custody transfer commissioning b. Close GC / start Windows Explorer (Close the GC program) c. Restarting the device d. Download ERRORLOG.TXT from the measuring element. Manual GC 9300 EN14 July, 24th 2018...
Bake out the measuring element Note These functions a. to g. have major effects on the functionality and are thus reserved for RMG Service!! The following functions can be selected by the “normal” user and started by click- ing on the “Execute” button.
When a parameter in the right window is clicked on, a window for changing the parameter appears: This window contains fields for entry of variable values or selection fields to change modes, depending on the type of parameter. In the example above, it is Manual GC 9300 EN14 July, 24th 2018...
The desired data can be filtered as follows with the selection fields: Left selection field: • All All conducted analyses, including calibrating and reference gas • S1 Gas Stream 1 Manual GC 9300 EN14 July, 24th 2018...
A calendar field appears in which the day of the archived data can be selected. Then you can jump back one, 10 or 100 entries with the buttons -100, -10, -1, +1, +10 and +100. 5.1.11 Log: Display of the log books Manual GC 9300 EN14 July, 24th 2018...
DSfG - Archive: Display of the DSfG archive The individual DSfG archive groups can be viewed on the “DSfG - Archive” screen. The archive groups exist from Stream1 – Stream4. 5.1.13 DSfG - Delete: Deletion of the DSfG archive Manual GC 9300 EN14 July, 24th 2018...
"User" appears. Click on the word "User", then on "Codeword" in the window on the right. Enter the user code using the displayed keyboard. III. Confirm your input by clicking the "OK" button. Manual GC 9300 EN14 July, 24th 2018...
STOP Measurement and calibration operation are disabled. BASIC-CAL Basic calibration for startup and for RMG service. Basic cali- bration may not be carried out during normal analysis opera- tion. NORM-CAL Manual calibration, which can be started at any time and has the same effect as automatic calibration and also takes 30 minutes.
Analysis". The measuring gas pressure is displayed on the right in the first line. In the window on the left, select "Measuring Element/Carrier Gas-I". The carrier gas pressure is displayed on the right in the first line. Manual GC 9300 EN14 July, 24th 2018...
Page 57
In the window on the left, select "Measuring Element/Channel-I". The III. pressure and temperature for column 1 are displayed in the window on the right. In "Measuring Element/Channel-2" and "Measuring Element/Channel-3" you can find the values for other columns. Manual GC 9300 EN14 July, 24th 2018...
After a longer service interval or if there is a possibility that impurities from unde- fined gas mixtures may have entered the measuring element, it is recommended that the measuring element is baked-out in order to remove these impurities. Note If necessary, consult RMG's service department before starting a heating process. 5.3.1 Bake-out process As of version V1.610-081, there is a new chapter 14.10 Bake-out under the chap-...
Page 59
• The operating mode must be set to STOP. • The carrier gas pressure must be correct. (For PGC9304 carrier gas 1 and -2.) • The calibration switch must be open in order to start the service function. Manual GC 9300 EN14 July, 24th 2018...
Page 60
• Remaining time of the heating process • Progress bar • Status display ("Bake out", "Cool down" and "Equilibrate") The heating process starts as soon as "Start bake-out process" has been confirmed twice. Manual GC 9300 EN14 July, 24th 2018...
Page 61
(always 1 hour) has to be waited. After cooling has been completed, the dialog box can be left again, regardless of whether the heating process was interrupted early or whether the heating process was successfully completed. Manual GC 9300 EN14 July, 24th 2018...
Page 62
5 Operation All actions of the heating process were logged into the event log (the timestamps do not correspond to a real process). Manual GC 9300 EN14 July, 24th 2018...
Data is read out and viewed from the following window: Reliable Measurement of Gas 5.5 Operation with RMGView You may find all required information in the manual of RMGView . This manual can be downloaded from the RMG internet homepage: www.rmg.com Manual GC 9300 EN14 July, 24th 2018...
Ana. & Cal.: Ethane content outside of user 7033 E Ethane Min/Max limits Ana. & Cal.: Propane content outside of user 7034 E Propane Min/Max limits 7035 E iso-Butane Min/Max Ana. & Cal.: i-Butane content outside of user Manual GC 9300 EN14 July, 24th 2018...
Page 65
Alarm limits for Wago analog output 13 violated 7073 E Wago AO14 alarm limits Alarm limits for Wago analog output 14 violated 7074 E Wago AO15 alarm limits Alarm limits for Wago analog output 15 violated Manual GC 9300 EN14 July, 24th 2018...
Page 66
7099 E Reserve §Min/Max (reserve for future functions) Calibration: At least one retention time outside Cal: retentions time of permitted limits Calibration: At least one response factor outside Response factor of permitted limits Manual GC 9300 EN14 July, 24th 2018...
Page 67
7152 W Temp.-1 warning limits Room temperature measurement error 7153 W Temp.-2 warning limits (reserve for future functions) outp current-1 warning Current output 1: Measured value outside of 7154 W limits limits Manual GC 9300 EN14 July, 24th 2018...
Page 68
Analysis: i-Butane content outside of permitted 7185 W Ana: iso-Butane §Min/Max limits Analysis: n-Butane content outside of permitted 7186 W Ana: n-Butane §Min/Max limits Ana: neo-Pentane Analysis: Neopentane content outside of per- 7187 W §Min/Max mitted limits Manual GC 9300 EN14 July, 24th 2018...
Page 69
7211 W Wago AO8 warning limits Warning limits for Wago analog output 9 violat- 7212 W Wago AO9 warning limits 7213 W Wago AO10 warning limits Warning limits for Wago analog output 10 vio- Manual GC 9300 EN14 July, 24th 2018...
Page 70
Measuring element is presently initializing 7243 H Analyser Flushing Measuring element is presently flushing Analyser calibrating Measuring element is presently calibrating Error with getting the chromatogram from the 7250 H Chromatogram (FTP) measuring element Manual GC 9300 EN14 July, 24th 2018...
Page 71
7319 W Wago AI10 warning limits Warning limits for Wago analog input 11 violat- 7320 W Wago AI11 warning limits Warning limits for Wago analog input 12 violat- 7321 W Wago AI12 warning limits Manual GC 9300 EN14 July, 24th 2018...
Page 72
Ana. & Cal.: n-Nonane content outside of user 7354 W n-Nonane Min/Max limits Ana. & Cal.: Oxygen content outside of user 7355 W Oxygen Min/Max limits 7356 W Helium Min/Max Ana. & Cal.: Helium content outside of user Manual GC 9300 EN14 July, 24th 2018...
Page 73
− For the current outputs, you can specify whether they are to be used for custody transfer (alarm) or not (warning). − For the limit values there are both warning limits (can be set freely via user code) and alarm limits (can only be set via calibration switch). Manual GC 9300 EN14 July, 24th 2018...
Embedded PC CPU ARM1176 Display 533 MHz 128 MB RAM 64 MB Flash Memory SD card 4 GB Operating system Windows CE 6.0 Digital inputs Number 13 mA 10 Hz Surge protection 6.8 V Manual GC 9300 EN14 July, 24th 2018...
Page 75
Load 700 ohms Surge protection 33 V Data interfaces Ethernet (2x) LAN 1 RMG network, measuring element connection DHCP server, DHCP client or fixed IP address LAN 2 Operator network DHCP client or fixed IP address USB (2x) Front For mouse, ext. hard disk or keyboard...
Appendix A: List of parameters The "Parameter list" is in a separate part of the manual called: PGC9300_AC_Parameter_en_12 This section of the manual can be accessed and downloaded – if required – from the homepage: www.rmg.com Manual GC 9300 EN14 July, 24th 2018...
Page 78
Contact Subject to technical modifications For further information If you would like to learn more about the products and solutions from RMG, please visit our web- site: www.rmg.com or your local support office or contact your local sales representative RMG Messtechnik GmbH Otto-Hahn-Straße 5...
Need help?
Do you have a question about the GC 9300 and is the answer not in the manual?
Questions and answers