Page 1
Model: Tachihi bldg, No.2, 6-1-3 Sakaecho, Tachikawa-shi, Tokyo, 190-8586 Japan Phone:+81-42-534-1413 Facsimile:+81-42-534-1438 YOKOGAWA EUROPE B. V. (THE NETHERLANDS) Euroweg 2, 3825HD, Amersfoot, THE NETHERLANDS Phone: 31-88-4641000, Facsimile: 31-88-4641111 Printed in JAPAN 2012/05 Compliance with the Radio Waves Act (Republic of Korea) This product complies with the Radio Waves Act (Republic of Korea).
• After reading this manual, please retain it for future reference. • YOKOGAWA is not liable for any incidental or consequential damages or losses including losses of data or chances of measurement, arising from accident, misuse or abnormal conditions of operation or handling.
Yokogawa is by no means liable for any damage resulting from use of the instrument in contradiction to these cautionary notes. This symbol signifies that improper usage may !...
Safe Usage –– Warnings & Cautions on Environment and Usage –– Warning –– Caution Environmental Warnings / Cautions K E E P T H E T H E R M O M E T E R KEEP THE THERMOMETER AWAY FROM DROPPING WATER AWAY FROM DIRECT SUNLIGHT, AND DO NOT USE IN WATER.
Specifications Model 53008-E Measurement range 0 to 500 ˚C Display range –10 to 650 ˚C φ 25 / 300mm ( D : S = 15 : 1 ) Field of view Optics mirror / silicon filter Sensing element Thermopile Spectral response 8 to 14 µ...
Name of Components * There is no button to turn ON/OFF the power supply. Pressing MEASURE button wakes up the thermometer. The power supply turns off automatically after 15 seconds' absence of pressing any button.
Page 7
Satety Precautions for Laser Products This Instrument uses a laser light source. This instrument is a Class 2 laser prod- uct as defined by IEC60825-1 Safety of Laser Products-Part1: Equipment classi- fication and requirements. In addition, this instrument complies with 21 CFR 1040.10 except for deviations pursuant to Laser Notice No.50, dated June 24, 2007.
Operation Set the battery supplied as an accessory and operate the unit according to the following procedure: ¡ ¡ ¡ ¡ ¡ Normal Measurement [Starting Normal Measurement] qPress MEASURE button to turn on the power supply. A laser beam is emitted and the measurement starts. (In the Normal Mea- surement mode, the unit takes measurements while MEASURE button is pressed.) wPoint the laser beam at a measuring object and aim it at the...
Field of View For the non-contact thermometer (infrared thermometer), the field of view (spot size) is specified depending on the distance from the thermometer to the measuring object as shown below. The temperature value displayed is the average temperature within the spot size. To take an accurate measurement, check the correlation between the size of object and the distance to it.
Memory Function 53008 can store 99-point temperature data in memory. Operate according to the following procedure: [Recording / MEM. mode] qPress and hold MEM./� button for two seconds or more to record the tem- perature value displayed at the time when you start pressing the button. Memory function is available in Normal Measurement mode, Continuous Measurement mode or HOLD mode (refer to page 8).
Settings This section gives you an explanation on how to set the following functions: "High Limit Temperature for Alarm", "Low Limit Temperature for Alarm", "Emissivity Ratio (DARK/BRIGHT mode)", "˚C mode" and "Recorded Erasing" "HOLD mode Selections". [Selecting Functions] qPress and hold SET/ENTER button for one second or more in the HOLD Mode (refer to page 8), and "SET"...
Setting –– "High Limit Temperature for Alarm" When the measured temperature exceeds the High Limit value, the alarm LED blinks in red and buzzer (high tone) sounds. [Setting "High Limit Temperature for Alarm"] qEnter the setting mode for "High Limit Temperature for Alarm" according to the procedure described in page 11.
Setting –– "Low Limit Temperature for Alarm" When the measured temperature is below the Low Limit value, the alarm LED blinks in green and buzzer (low tone) sounds. [Setting "Low Limit Temperature for Alarm"] qEnter the setting mode for "Low Limit Temperature for Alarm" according to the procedure described in page 11.
* If you wish to cancel the Record Erasing, press � button to select "ESC" sign on the display and press SET/ENTER button. Then, you can return to the setting mode again. * Please note that Yokogawa M&C is not liable for any loss of data. [Record Erasing Cancel mode]...
Setting –– "HOLD made Selections" [Setting "HOLD mode Selections"] qEnter the setting mode for "HOLD mode Selections" according to the procedure described in page 11. wPress � or � button in the setting mode to select ON or OFF. The initial setting is OFF. (NOR HOLD) ePress SET/ENTER button to set ON or OFF.
Maintenance / Battery Maintenance [Filter] Dust, stain or scratch on the filter causes incorrect measurement. If the filter is stained, clean the filter with a lens-cleaning blower or wash it with clean water. If the filter is still stained, gently wipe the stain off with a cotton swab or lens cleaning cloth, which may be moistened with ethyl alcohol.