ETCR 016 User Manual

High accuracy clamp leakage sensor

Advertisement

ETCR016 High Accuracy Clamp Leakage Sensor
Thanks for your purchase of ETCR016High Accuracy Clamp Leakage
Sensor of our company. For better use of the product, please make sure:
---To read this user manual in details.
---To abide by the safety regulations and precautions strictly.
 Under any circumstance, it shall pay special attention on safety in use of this sensor.
 Pay attention to words and symbols stick on the panel.
 Keep the pliers clean, maintenance regularly.
 Stop using the sensor when there is a rupture or break.
Please don't keep or store the sensor in the spot with high-temperature and moisture, or condensation,
and under direct daylight radiation for a long time.
 This sensor is only to be used, disassembled, and repaired by qualified personnel with authorization.
 When it may cause hazard by continuous use for the reason of the sensor itself, it shall immediately stop
using it and deposit it at once, leaving it for disposal by authorized agency.
 For risk of danger icon in manual
manual content.

I. Introduction

ETCR016High Accuracy Clamp AC Leakage Sensor use double shield technology, with small size, high
accuracy, good stability, strong anti-interference ability characteristics.
ETCR016 High Accuracy Clamp AC Leakage Sensor is widely applied in AC current leakage, current,
power and energy measurement with high precision and small phase error of electricity, communications,
meteorology, railway, oilfield, construction, measurement, scientific and research teaching unit, industrial and
mining enterprises, which can be connected to a variety of high precision digital multi-meter and data recorder.

II. Technical Specifications

Function
Range
Resolution
Accuracy
Coils Turn
Phase Error
Reference Load
Weight
Dimension
Current transformer
Clamp Size
Detection method
Dimension
Out put
Output Interface
Output line
Electric Field
Interference
Current Frequency
Frequency

User Manual

, users must perform safety operations strictly in compliance with the
AC leakage current, current, high harmonic current, phase, power, power,
power factor detection
0.000mA~60.00A AC
0.001mA AC
± 1.0% FS (50Hz / 60Hz; 23
center of the jaw)
800:1
≤2°(50Hz/60Hz; 23
±2
RL: 0-600mA≤300ohm; 0-6A≤30ohm; 0-60A≤3ohm;
138g
110×53.4×22 mm
Around 1A/1mA
Φ16mm
Clamp CT
115mm×70mm×33mm
Current sensing output
3.5mm audio plug(tail plug, banana pin, or BNC pin)
2-core shielded wire, wire length 2m
About 5mA when the external electric field 100A, 10mm nearby
45H-60Hz(when measuring big current)
10Hz-100kHz
± 2
, 70% RH or less, the wire is in the
)
- 1 -

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 016 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ETCR 016

  • Page 1: User Manual

    ETCR016 High Accuracy Clamp Leakage Sensor User Manual Thanks for your purchase of ETCR016High Accuracy Clamp Leakage Sensor of our company. For better use of the product, please make sure: ---To read this user manual in details. ---To abide by the safety regulations and precautions strictly. ...
  • Page 2 3 wires) Clamp four wires together to measure the leakage current of three phase four wires.(Note: 4 wires) Manufactured by ETCR Electronic Technology Company Address: F-3F, No.4 Pengshang Zhifu Road, Jiahe, Baiyun District, Guangzhou, Guangdong, China Post Code: 510440 Tel: (86-20)62199556...

Table of Contents