Dickson KT800 Getting Started

Remote sensing temperature recorder

Advertisement

Default Recorder Settings
DICKSON KT800/802/803/804/856
7-day chart rotation
0 to 100°F (KT800/02/03/09/56)
0 to 1000°F (KT804)
Quick Start
1. Your recorder has been preset to operate using the
Quick Start
most popular settings.
1. Remove the protective pen caps.
KT800/02/03/56
Plug in thermocouple probe(s).
Recording Time: 7 day
Temperature Range: 0 to +100°F
Note: Be sure to observe proper polarity when
KT804
connecting wires from transmitter. (See figure 3)
Recording Time: 7 day
The KT809 features an external K-TC connector.
Connect the thermocouple probe provided to the
Temperature Range: 0 to +1000°F
external K-TC connector. Use the waterproof tape
To change the range and recording time, please
provided to wrap the connector and jack to create a
reference Dip Switch Set-Up in this manual.
watertight seal.
2. Remove the protective pen caps.
2. I nsert batteries and plug in AC adapter. (See figure 3.)
3. Plug in thermocouple probe(s).
(KT800/09) requires 4 AA batteries only.) The unit
NOTE: Be sure to observe proper polarity when
will power on.
connecting wires from transmitter. (see figure 3)
3. T he instrument will move pen(s) to the appropriate
4. Insert batteries and plug in AC adapter. (see figure
readings.
3) (KT800 require 4 AA batteries only) Unit will
4. I nstall the chart that matches the Dip Switch settings
power on.
P ress the Pen Home key to move the pens to the
5. The instrument will move pens to the appropriate
outside of the chart. The pen is automatically raised
readings.
off the chart.
6. Install the chart that matches the Dip Switch settings
you have selected.
Remove the old chart, place the new chart on the
Chart Hub being certain that the edge of the chart
a. Press the PEN HOME key to move the pens to the
slides under the Chart Guide Clips located at the
outside of the chart. The pen is automatically
outside of the chart.
raised off the chart.
5. S et the appropriate time. There are two ways to
b. Remove the old chart, place the new chart on
adjust the chart and set the appropriate time:
the Chart Hub – being certain that the edge of the
chart slides under the Chart Guide Clips located at
a. S et the chart time manually by inserting a coin into
the outside of the chart.
the groove in the chart hub and turning clockwise
until the correct hour (and day if applicable) on the
7. Set the appropriate time. There are two ways to
chart is referenced to the timing arrow (just to the
adjust the chart and set the appropriate time:
right of the pen tip on the dial).
a. Set the chart time manually by inserting a coin
Display Symbols KT8
Display Symbols (optional)
Figure 1
Low battery
indicator
Alarm
Enabled
Thermocouple
Type
Button Operations
www.DicksonData.com • 1-800-323-2448 • Fax 1-800-676-0498
ON/OFF
Getting Started
Remote Sensing Temperature Recorder
-
+
K
KT800/802/803/804/809/856
The following feature should be used for fine
adjustments only.
b. T o adjust the chart time, press and hold
Adjust-Up and Adjust-Down buttons located
on the back of the unit next to the dip switch (see
into the groove in the chart hub and turning
Figure 3).
clockwise until the correct hour (and day if
Note: Model KT809 only, the Adjust-Up and
applicable) on the chart is referenced to the
Adjust-Down buttons are located under the
timing arrow (just to the right of the pen tip on
battery door on the back of the unit.
the dial).
The green LED will blink rapidly for about five
b. (This feature should be used for fine adjustments
seconds, then the LED will remain solid green.
only) To adjust the chart time, press and hold
While in this state the Adjust-Up button will move
ADJUST-UP and ADJUST-DOWN buttons located
the chart backward (counter clockwise) and the
on the back of the unit next to the dip switch (see
Adjust-Down button will move the chart forward
figure 3). The green LED will blink rapidly for
(clockwise).
about five seconds, then the LED will remain solid
green. While in this state ADJUST-UP button will
Rotate the chart until the correct hour (and day if
applicable) on the chart is referenced to the timing
move the chart backward (counter clockwise)
arrow. Once you have set your chart, press the Pen
and ADJUST-DOWN button will move the chart
Home button to exit Chart Adjust Mode. The unit will
forward (clockwise). Rotate the chart until the
take one minute to exit the Chart Adjust Mode once
correct hour (and day if applicable) on the chart is
Pen Home is pressed.
referenced to the timing arrow. Once you have
set your chart, press the PEN HOME button to exit
P ress Pen Home to move the pens back onto
Chart Adjust Mode. The unit will take one minute
the chart.
to exit Chart Adjust Mode once PEN HOME is
6. P lace the KT8 recorder on a flat vibration-free
pressed.
surface. Be sure it is in a vertical position and level.
8. Press PEN HOME to move the pens back onto the
For best performance and longevity, the location
chart.
should be a clean environment, free from dust
9. Place the KT8 recorder on a flat vibration-free
and corrosive fumes. Do not exceed temperature
surface. Be sure it is in a vertical position and level.
specifications.
For best performance and longevity, the location
Wall Mount Keyhole slots are provided on the KT8
should be a clean environment, free from dust
recorder for wall mounting.
and corrosive fumes. Do not exceed temperature
specifications.
Wall Mount: Keyhole slots are provided on the KT8
recorder for wall mounting.
MAX
F or C
MIN
position. If the pen(s) are located on the chart, press the
PEN HOME key to move the pen(s) to the outside edge
Remote Sensing
Temperature Recorder
Current Reading
or
Flashing "ALRN" when
in Alarm Condition
For setting
Alarms only

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the KT800 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Dickson KT800

  • Page 1 Pen Home is pressed. referenced to the timing arrow. Once you have 3) (KT800 require 4 AA batteries only) Unit will 4. I nstall the chart that matches the Dip Switch settings set your chart, press the PEN HOME button to exit P ress Pen Home to move the pens back onto power on.
  • Page 2: Button Operations

    KT800/802/803/804/809/856 Remote Sensing Temperature Recorder Button Operations Figure 2 Figure 3 Adjust Buttons Dip Switches KT803/04/56 Relay Contacts (KT856) Pen Power Home Alarm On/Off Battery Compartment KT800/02/09 Adapter Power Jack Pen Home Figure 3–KT809 Battery Case & Switch Cover 5 screws On/Off On/Off The On/Off key turns the unit on and off. Pen Home (Down Arrow) If the pens are located on the outside edge of the chart, press the Pen Home to move pen(s) to recording position. If the pens are located on the chart, press the Pen Home key to move the pen(s) to the outside edge of the chart. Adjust Alarm Optional...
  • Page 3: Alarm Relays

    0 to 150° F/C 05/55/56 scroll through the five options until the alarm button Remember to install the correct chart to match corresponding switch setting is pressed to exit from alarm adjust. Each press of the Pen Home or Alarm button will store the new DICKSON • Addison, IL www.DicksonData.com • settings. The unit will take one minute to exit Alarm Set Mode once the Alarm button is pressed. KT8 Dip Switch Setting Up=On Down=Off 4. I f the alarm is triggered, the LED will show as solid...
  • Page 4: Led Indicators

    #5 up Note: (Unit should not be placed in an environment #6 down below 32°F) LED Indicators KT804 Dip Switches for Ranges • AC Power with battery backup–Solid green #4 down • A C Power with low battery or no battery–Blinks red #5 up -20 to +20°F/C • Battery only–Blinks green #6 down • Battery only (Low battery) –Solid red #4 down Note: KT800/09 operates on battery only and will blink one time every minute to conserve battery life. +20 to +100°F/C #5 up #6 up CALIBRATION #4 up Your instrument was carefully tested and calibrated before +50 to +150°F/C #5 down being shipped from the factory. For greatest accuracy, we recommend factory re-calibration every 6–12 months. Call #6 down customer service at (630) 543-3747. #4 down If you wish to do calibration yourself, follow these #5 down 0 to +1000°F/C...
  • Page 5: Need To Know

    Each time the The Door and Battery Cover gaskets should maintain the Pen Home or On/Off button is pressed, the adjustment IP66 rating for 1 year under normal use. To replace gaskets, currently in progress will be stored. The unit will take one contact Dickson Customer Service. minute to exit Pen Adjust Mode once the On/Off button is pressed. KT800/09 Pen Movement TROUBLESHOOTING The pen movement for the KT800/09 is slower than on the AC powered models in order to conserve battery power.
  • Page 6: Warranty

    Why won’t the unit power up? • Pen(s) not inserted on pen arm(s) all the way • R emove the batteries and power adapter for a • T o adjust pen to match chart see Pen Adjust in the minute or two, this will reset the unit. The unit should Need to Know section of this manual. Note: KT800 respond when the adapter is plugged in. does not have a display. Water is getting inside the case. Why does the display read PROB? • C heck the battery cover to make sure it is screwed • C heck the K-TC probe. Make sure it is properly...

This manual is also suitable for:

Kt809Kt856Kt803Kt802Kt804

Table of Contents