Settings; Selecting Concentration Units - jenway 7315 Operating Manual

Spectrophotometers
Table of Contents

Advertisement

pp m
0
6.2.2

Settings

The settings menu enables the wavelength, units, resolution, standard or factor to be set and is
accessed from the expanded operating menu by pressing the key adjacent to the settings icon. Once
all of the required settings have been entered press the key adjacent to the tick icon to save and return
to the expanded operating menu.
Selecting resolution
Selecting
concentration units
When setting the method parameters either the standard or the factor should be selected. The standard
should be used if the factor is not known as selecting this option will calculate the factor. If the factor is
known it is not necessary to measure a known standard's concentration. When the standard or factor is
not selected the value should be set to 1.00.
6.2.2.1

Selecting Concentration Units

The units of concentration can be selected from a number of options: no units, %, ppm, EBC, SRM,
mEq/l, mEq, M, mM, µM, nM, U, U/l, U/ml, g/l, mg/l, µg/l, ng/l, g/dl, mg/dl, µg/dl, mg/ml, µg/ml, ng/ml,
µg/µl, ng/µl, mol/l, mmol/l.
----
mEq
U/ml g/l
mg/dl
µg/µl
28
2.008
1.000
400nm
4
0
0
ppm
Fig 6.2.2.1 – Settings Menu
%
ppm
EBC
SRM mEq/l
M
mM
µM
nM
U
U/l
mg/l
µg/l
ng/l
g/dl
µg/dl
mg/ml µg/ml ng/ml
ng/µl
mol/l
mmol/l.
The settings menu is accessed through the expanded
operating menu by pressing the key adjacent to the
settings icon. In the settings menu press the key below
the wavelength icon.
This will open a number entry screen. Use the keys at the
bottom of the screen to select the digit to be adjusted.
Use the keys adjacent to the arrow icons to increase or
decrease the wavelength to the required number. Press
the key adjacent to the tick icon to save the changes and
return to the settings menu.
2.008
1.000
400nm
In the settings menu press the key below the units icon.
This opens the unit selection screen which displays all
the different units. Use the keys adjacent to the arrow
icons to navigate around the screen to select the required
units. Once the required units have been highlighted
press the key adjacent to the tick icon to save and return
to the settings menu. The selected unit will be displayed
in the minimal and expanded operating menu along with
absorbance and selected wavelength.
Tick
Standard menu
Factor menu
Selecting a
wavelength