ELECTROMAGNETIC COMPATIBILITY

GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS
The NICE1 is intended for use in the electromagnetic environment specified below. The customer
or the user of the NICE1 should assure that it is used in such an environment.
EMISSIONS TEST
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The NICE1 is intended for use in the electromagnetic environment specified below. The customer
or the user of the NICE1 should assure that it is used in such an environment.
IMMUNITY TEST
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
COMPLIANCE
Group 1
Class B
Class A
Complies
IEC 60601
TEST LEVEL
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT) for 5 s
N/A
ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
The NICE1 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
The NICE1 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
COMPLIANCE LEVEL
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV differential mode
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT) for 5 s
N/A
ELECTROMAGNETIC
ENVIRONMENT – GUID-
ANCE
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material,
the relative humidity
should be
at least 30 %.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the NICE1
requires continued
operation during power
mains interruptions, it is
recommended that the
NICE1 be powered from
an uninterruptible power
supply or a battery.
NICE1 does not contain
magnetically sensitive
components
NICERECOVERY.COM