TROUBLESHOOTING & PUMP ALARMS STEP BY STEP GUIDE Why your pump alarms and how to stop it Read the screen Press the START/STOP button to silence the alarm See overleaf for a list of reasons why your pump is alarming and how to deal with each alarm...
Page 2
ALARM REASON SOLUTION The set is not fitted or Reinsert the giving set as per instructions is fitted incorrectly NO SET The pressure sensor Clean the sensors and reinsert cassette is dirty Make sure the door is closed Tape door to pump tightly to continue Door is broken, cracked DOOR feeding and call your Nutricia Homeward...
Page 3
ALARM REASON SOLUTION Press and hold the ON/OFF button until it beeps and is powered off END OF The pump has If the END OF DOSE alarms DOSE administered the dose inappropriately contact your Nutricia Homeward Enteral Nurse Specialist or the Out of Hours Advice Line Make sure there are no kinks in The pump detected an...
Page 4
HOW TO STOP THE OCC ALARM Press the START/STOP button Detach the giving set Solve the occlusion problem (see below) Re-insert the giving set and close door Press the START/STOP button THE MOST COMMON SET OCCLUSION TUBE KINKING: Narrow the tube, release kinking...
Page 5
ALARM REASON SOLUTION The battery capacity is Connect the pump to the running low. There is electrical supply and charge BATT approximately one hour of for 6 hours. The pump can be feeding left when a rate of used while connected to the 125ml/hour is set electrical supply Contact your Nutricia...