Setting Clinician Values; Protocols; Clinician Value Details; Rate - PRIZM Micro-Z MINI Operation Manual

Pulsed dc current tens stimulator
Table of Contents

Advertisement

5.0 SETTING CLINICIAN VALUES

5.1 Protocols

The Micro-Z Mini™ is factory programmed for
two treatment protocols.
P1: Protocol One, 30 Minute Treatment,
is a two-segment 30-minute treatment: Segment
One runs 100 pps (Hz) for 15 minutes, Segment
2 runs 10 pps (Hz) for 15 minutes.
P2: Protocol Two, 8-Hour Treatment,
is an 8-hour routine that consists of 20 minutes
of therapy followed by 40 minutes of rest every
hour for 8 hours. The 20-minute treatment
consists of 10 minutes at 80 pps (Hz) and 10
minutes at 8 pps (Hz).
After treatment P1 or P2 is complete, the Micro
Z Mini™ will automatically turn itself off. Every
10 seconds during the protocol run the display
will show time remaining in hours and minutes or
minutes and seconds. During the Dwell time of
protocol 2 (P2) the intensity setting will fl ash.

5.2 Clinician Value Details

The Micro-Z Mini™ can be programmed for
8
many different treatment protocols with the Prizm
Medical Micro-Z Mini™ docking station to a PC.
Check with your clinician or dealer for these
options.
The Micro-Z Mini™ delivers a high
voltage twin peak monophasic wave
form. The maximum voltage as
indicated on the packaging would
coincide with the numeric intensity
value on the LCD display
Contact Prizm Medical, Inc.
for additional information including
ranges, accuracy and precision
of the displayed intensity readings.

5.3 Rate

Rate selects the number of times the waveform
repeats every second. Note that P1 is
programmed at 100 pps for 15 minutes and

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for PRIZM Micro-Z MINI

Table of Contents