NOTE:
The EverFlo has Essential Performance as defined in ISO 80601-2-69. The EverFlo will deliver oxygen in
both normal and single fault conditions per the specifications in this manual, or in the case of a power
supply failure, low oxygen concentration, or device malfunction, an alarm condition will occur.
Electrical,
AC Power Consumption
Models 1020000, 1020001
1020002, 1020003
1020002BR, 1020003BR
1020014, 1020015
1039362, 1039363
1039364, 1039365
1020004, 1020005 220 VAC ±10%, 320 W, 60 Hz
1020006,1020007, 1020008
1020009, 1020010, 1020011
1020012, 1020016, 1020017
1020020, 1039366, 1039367
1039368, 1039370, 1104000
Oxygen
Oxygen Concentration*
(All Models except
as noted below)
Models 1020007, 1020008
1039367, 1039368, 1104000
* Device operation above or outside of the voltage, LPM, temperature, humidity and/or altitude values specified may
decrease oxygen concentration levels.
** Oxygen flow rate <1 LPM requires low range flow meter accessory.
Maximum output pressure limited to 6.5 PSIG (44.8 kPa).
120 VAC ±10%, 350 W, 60 Hz
230 VAC ±10%, <300 W, 50 Hz
1125558
1102443 230 VAC ±10%, <300 W, 60 Hz
1020013 220 VAC ±10%, <300W, 60 Hz
90–96% from 1 to 5 LPM**
87–96% from 1 to 5 LPM**
EverFlo / EverFlo Q User Manual
15