Download Print this page

T&D TR-71nw Introductory Manual

Tr-7nw series thermo recorder

Advertisement

Quick Links

Thermo Recorder
TR-7nw Series Introductory Manual
What can the TR-7nw Series devices do?
Via Cloud
Auto-upload of Recorded Data
*1
Wired LAN Communication
Apply Settings
TR-7nw Series
*1: To auto upload recorded data to the "T&D WebStorage Service", it is necessary to have Internet access via wired LAN and a T&D WebStorage Service User ID.
The "T&D Data Server" application can be used to auto upload data to your own PC rather than "T&D WebStorage Service". (See
Via USB
USB Communication
TR-7nw Series
Common Operations
Setting up the Device
11
Connect sensor(s) to the
device.
Image of TR-71nw
To ensure a proper connection, make sure that the sensor connector is completely
inserted.
21
Remove the battery cover and insert
batteries.
• Make sure not to mistake + / -.
• Always use new batteries.
31
The measurement will be displayed on the LCD screen.
During Sensor Error (no sensor connected /
Under Normal Conditions:
improper connection / damaged):
Starting Recording
• Upon the start of a new recording session all stored data will be deleted from
the device.
• Recording Interval settings cannot be changed while a recording session is in
progress. Stop recording first and then change settings.
• During recording, the power cannot be turned off.
11
Click the <INTERVAL> button to
display the current recording interval.
21
Press the <INTERVAL> button until the display flashes.
© Copyright T&D Corporation. All rights reserved.
2016.01 16504980016 5th Edition
http://www.tandd.com/
Graph Display
Download Data
Setup
View with a Web Browser
printed on the back of this paper.)
Setup and Data Download
PC
Graph Display
31
With each pressing of the <INTERVAL> button the recording
interval time will change. When the desired interval appears, stop
pressing the button.
Within a few seconds, the display will return to measurement
mode to indicate the recording interval is set.
41
Start recording by pressing
and holding the <REC/STOP>
button until the
mark
appears in the display.
Below are some examples of recording intervals and maximum recording times1
1 sec.
30 sec.
5 min.
15 min.
About 2 hours
About 2 days
About 27 days
About 83 days
In order to stop a recording session, press and hold the <REC/STOP> button until the
mark disappears from the display.
Details about the LCD Display
Other than the measurement display, the following is displayed on
the LCD monitor.
Data Transmission in Progress
Internet Connection Status
ON : Connected to the Internet.
BLINKING : Unable to connect to the Internet.
OFF : Auto-upload is set to "OFF".
Displayed Measurement Channel
LAN or USB Communication in Progress
Recording Status
ON: Recording in progress
BLINKING: Waiting for programmed start
OFF : Recording stopped
Recording Mode
ENDLESS : Upon reaching the logging capacity of 8,000 readings, the oldest data is
overwritten and recording continues.
ONETIME : Upon reaching the logging capacity of 8,000 readings, [FULL] appears on
the LCD display and recording automatically stops.
* Use "TR-7wf/nw for Windows" or "T&D Thermo" to change the recording mode.
(Cannot be changed by buttons on TR-7nw.)
Amount of Recorded Data in Device
Low Battery Level
* Please change the batteries as soon as you see this mark. If not, the device enters
sleep mode ([SLP] appears on LCD) and stops all operations. If the batteries are further
left unchanged, all previously recorded data will be lost.
Indicates Auto-upload Interval when using <INTERVAL> button
Indicates Recording Interval when using <INTERVAL> button
Operations via Cloud
Making Auto-Upload Settings
By making settings for auto-upload, it is possible to automatically
upload recorded data to the online "T&D WebStorage Service".
In the auto-upload, the maximum number of readings which can be stored in
"T&D WebStorage Service" is 80,000. When this limit is exceeded, the oldest data
will be overwritten and deleted.
11
Connect a LAN cable to the Ethernet port.
If you wish to use a static IP address instead of DHCP, change the network settings on
our software "TR-7wf/nw for Windows" by connecting the device to a PC via USB. For
details about operations please see the "Help" menu in the software.
21
Click the <INTERVAL> button
twice to display the current
upload interval.
31
Press the <INTERVAL> button until the display flashes.
41
With each pressing of the <INTERVAL> button the upload interval
time will change. When the desired interval appears, stop
pressing the button.
If you wish NOT to use the auto-upload feature, select "OFF".
Within a few seconds, the display will return to measurement
mode to indicate the upload interval is set.
Below are some examples of upload intervals and the corresponding estimated
battery life1
In general, the shorter the upload interval, the shorter the battery life.
Upload Interval
1 min.
10 min.
Battery Life
About 10 days
About 2 months
• All estimates are based on operations carried out with a new battery and are in no way a
guarantee of actual battery life.
• The auto-upload interval can also be changed during a recording session.
Installing the Device
Install the device to the actual place it will be used.
60 min.
About 333 days
Internet
Router / HUB
Checking Connection Status
By pressing the <REC/STOP> button, recorded data will be sent to "T&D WebStorage Service".
If the transmission is successful,
will still be ON even after communication is complete
and
turns OFF.
is blinking : Ethernet connection established and data is being transmitted.
is ON : Internet connection is successfully established.
is blinking : Internet connection failed. Please check the network settings.
Checking Recorded Data
11
Access "T&D WebStorage Service" from the web browser on
your PC or mobile device.
http://www1webstorage-service1com/
21
Follow the directions as they appear to get a User ID.
If you already have your User ID, go to the next step.
31
Login by entering the registered User ID and Password.
41
Clicking the [Account] button will take you to the Account
Management page.
Auto-upload settings enable this product to automatically upload data to T&D WebStorage Service.
Please read the T&D WebStorage Service License Agreement carefully before making settings.
* Making Auto-upload Settings indicates your acceptance of this Agreement.
51
Select [Add/Remove Devices] to open the registration page, and
enter the registration code for the TR-7nw Series device to be
added.
The registration code can be found on the supplied Registration
Code Label.
61
After data is sent to the "T&D WebStorage Service", either by
auto-upload or by pressing the <REC/STOP> button on the
device, the added TR-7nw Series device and its measurements
will appear in the [Current Readings] tab page.
If measurements are not displayed:
• Check the device screen to make sure
when recording is not in progress.
• Try sending data to "T&D WebStorage Service" again by pressing the <REC/STOP> button
on the device.
• See "Checking Connection Status" in step 2, and check if the device has a LAN connec-
tion.
If the display of measurements is not refreshed:
We recommend making the recording interval shorter than the auto-upload interval. If the
recording interval is set to longer than the upload interval, the same measurement is displayed
until the next recording occurs and the new data is uploaded.
If you cannot find the registration code:
1 hr.
12 hrs.
The registration code can also be found by using the software "TR-7wf/nw for Windows" via a
About 1 year
About 1.5 years
USB cable connected to a PC. See
Viewing Data from other than a Web Browser
Recorded data which was sent to the "T&D WebStorage Service"
can also be viewed from "T&D Thermo" and/or "T&D Graph".
See
T&D Thermo and
of this paper.
Check Internet Connection Status
Saving Auto-Uploaded Data
Recorded data which has been uploaded to the
"T&D WebStorage Service" can be saved to a PC by using
"T&D Graph" and accessing "T&D WebStorage Service".
See
T&D Graph printed on the back of this paper.
Changing the Settings
By using the "T&D WebStorage Service", it is possible to change the
following settings.
Device Information:
Device Name, Group Name, Recording Interval, Recording Mode,
Auto-upload Interval, Channel Name, Time Settings, Temperature Unit
Warning Settings:
Lower Limit, Upper Limit, Sensor Warning, Judgement Time
The changes in settings will be applied when the device communicates with
the "T&D WebStorage Service". To apply the changes immediately, press the
<REC/STOP> button on the device after settings are completed.
is ON. Recorded data will not be uploaded
Operations via USB
on the back of this paper.
T&D Graph printed on the back

Advertisement

loading

Summary of Contents for T&D TR-71nw

  • Page 1 Start recording by pressing guarantee of actual battery life. • The auto-upload interval can also be changed during a recording session. and holding the <REC/STOP> Image of TR-71nw Recorded data which was sent to the "T&D WebStorage Service" button until the mark can also be viewed from "T&D Thermo"...
  • Page 2 *5: Battery life varies depending upon multiple factors including frequency of communication, LAN environment, and internal sensor of the TR-71nw will report a higher temperature than actual; we ambient temperature, recording interval, and battery performance. All estimates are based on operations recommend using an external temperature sensor.

This manual is also suitable for:

Tr-72nwTr-72nw-h